Fluke-biomedical 451P User Manual Page 29

  • Download
  • Add to my manuals
  • Print
  • Page
    / 30
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 28
Ion Chamber Survey Meter
Maintenance
19
Replaceable Parts Information
Fluke maintains a complete inventory of all normal replaceable parts. To place an
order, or to obtain information concerning replaceable parts, contact Fluke. See
the How to Contact Fluke section in this manual.
Table 3. Replaceable Parts List
Part Number Description
3567752 Keypad Overlay, Red
3567765 Handle, Yellow
3567776 Handle, Foam Grip, Black
3463380 Check source, CS-137, 10 UCI, Flat disc, 1-inch diameter
614487 Battery, 9 V, Alkaline
Recalibration and Service Information
Fluke recommends a once yearly calibration on your Survey Meter. If your
Survey Meter needs recalibration or repair, we request that you contact Fluke.
See the How to Contact Fluke section in this manual.
More information concerning the operation, application, or service of your Survey
Meter can be obtained from the applications engineer at the numbers listed
above.
Troubleshooting
Caution
For safe operation and maintenance of the Product, use caution
when working on this Product. If the device indicates a
relatively high radiation rate, there is a possibility that the
Product has been contaminated. High rates can quickly cause
the operator’s cumulative exposure to increase with the
potential for injury.
Note
Install fresh batteries prior to performing any calibration on the
Survey Meter.
If the Survey Meter fails self-test, it will remain locked on the firmware revision
number. See the How to Contact Fluke section.
Page view 28
1 2 ... 24 25 26 27 28 29 30

Comments to this Manuals