Impulse 6000D Defibrillator Analyzer Impulse 7000DP Defibrillator/Transcutaneous Pacer Analyzer Users Manual August 2007, Rev. 1, 1/08 © 2007 Fl
7000DP, 6000D Users Manual iv Energy Measurement Test... 60 Cha
v List of Tables Table Title Page 1. Symbols...
7000DP, 6000D Users Manual vi 18. Pacer Demand Mode Test Custom Parameters ... 70
vii List of Figures Figure Title Page 1. Top-Panel Controls and Connections...
7000DP, 6000D Users Manual viii 18. ECG Main Menu ...
1 Defibrillator Analyzer Introduction The Impulse 6000D and 7000DP (hereafter the Analyzer) are portable, battery-powered precision instruments for t
Impulse 6000D, 7000DP Users Manual 2 Table 1. Symbols Symbol Description W Important information; refer to manual. ~ Do not dispose of this product a
Defibrillator/Transcutaneous Pacemaker Analyzer Safety Information 3 • Do not operate the Analyzer with the battery eliminator connected, unl
Impulse 6000D, 7000DP Users Manual 4 Instrument Familiarization Figure 1 and Table 2 describes the top-panel controls and connections of the Analyzer.
Defibrillator/Transcutaneous Pacemaker Analyzer Instrument Familiarization 5 Table 2. Top-Panel Controls and Connections Item Name Descriptio
Warranty and Product Support Fluke Biomedical warrants this instrument against defects in materials and workmanship for one year from the date of orig
Impulse 6000D, 7000DP Users Manual 6 Figure 2 and Table 3 describes the rear-panel connections of the Analyzer. FLUKE BIOMEDICAL6920 SEAWAY BLVDEVERET
Defibrillator/Transcutaneous Pacemaker Analyzer Turning the Analyzer On and Off 7 Table 3. Rear-Panel Connections Item Name Description 1 Ch
Impulse 6000D, 7000DP Users Manual 8 Accessing the Analyzer Tests The Analyzer uses a series of menus to access various Analyzer functions and setup v
Defibrillator/Transcutaneous Pacemaker Analyzer Analyzing Defibrillators 9 OFFDEFIBDefibrillator/PacerPACER++ fak11.eps Figure 6. Defibrillato
Impulse 6000D, 7000DP Users Manual 10 fak04.eps Figure 7. Defibrillator Energy Test If the waveform characteristics are correct, then charge the defi
Defibrillator/Transcutaneous Pacemaker Analyzer Analyzing Defibrillators 11 the time from the ECG heart beat ‘R’ wave to the onset of the defi
Impulse 6000D, 7000DP Users Manual 12 The Analyzer senses the discharge and the charge time appears in the display. Press the softkey labeled Measure
Defibrillator/Transcutaneous Pacemaker Analyzer Analyzing Pacemakers (7000DP only) 13 OFFDEFIBDefibrillator/Pacer7000DP Only50 ΩOnlyPACER++
Impulse 6000D, 7000DP Users Manual 14 Press the softkey labeled Brand to activate the brand list and scroll through the list using G or H. When the co
Defibrillator/Transcutaneous Pacemaker Analyzer Analyzing Pacemakers (7000DP only) 15 When testing the attached pacemaker, operating in the co
Notices All Rights Reserved © Copyright 2007, Fluke Biomedical. No part of this publication may be reproduced, transmitted, transcribed, stored in a r
Impulse 6000D, 7000DP Users Manual 16 Next, press the softkey labeled Demand. Pressing the softkey labeled Overdrive causes the Analyzer’s ECG signal
Defibrillator/Transcutaneous Pacemaker Analyzer Analyzing Pacemakers (7000DP only) 17 labeled Summary displays a summary of the test which can
Impulse 6000D, 7000DP Users Manual 18 Sensed Refractory Period (SRP) The Analyzer next generates a second simulated ECG signal immediately trailing th
Defibrillator/Transcutaneous Pacemaker Analyzer Simulating ECG Signals 19 Note If noise is present in the ECG with the battery charger plugged
Impulse 6000D, 7000DP Users Manual 20 ECG Monitor fak09.eps Figure 20. ECG Connections
Defibrillator/Transcutaneous Pacemaker Analyzer Simulating ECG Signals 21 The performance signal controls consist of a waveform selection, amp
Impulse 6000D, 7000DP Users Manual 22 Setting the Analyzer for an Interactive ECG/Pacer Asystole Mode Simulation From the ECG main menu, press the sof
Defibrillator/Transcutaneous Pacemaker Analyzer Simulating ECG Signals 23 amplitude, press the softkey labeled Amplitude and use G or H to scr
Impulse 6000D, 7000DP Users Manual 24 Performing a Noise Immunity Test This qualitative test verifies the pacemaker’s ability to filter line frequency
Defibrillator/Transcutaneous Pacemaker Analyzer Setting Analyzer Setup Functions 25 fak23.eps Figure 28. Battery Setup Screen While in the ba
Standard Terms and Conditions Refunds and Credits Please note that only serialized products and their accessory items (i.e., products and items bearin
Impulse 6000D, 7000DP Users Manual 26 Setting the Display Contrast The Analyzer’s display contrast can be set in one of two ways. First, when the Anal
Defibrillator/Transcutaneous Pacemaker Analyzer Maintenance 27 written test procedure), and integrates all test results into a single test rep
Impulse 6000D, 7000DP Users Manual 28 W Caution Do not use spray cleaners on the Analyzer; such action may force cleaning fluid into the Analyzer and
Defibrillator/Transcutaneous Pacemaker Analyzer Accessories 29 Accessories Table 4 lists the accessories for the Analyzer. Table 4. Accessorie
Impulse 6000D, 7000DP Users Manual 30 Specifications General Specifications Temperature Operating ...
Defibrillator/Transcutaneous Pacemaker Analyzer Specifications 31 Defibrillator Analyzer Specifications Energy Output Measurement Compatible D
Impulse 6000D, 7000DP Users Manual 32 Sample rate...250 kHz (4 μs sample) Maximum Average Pow
Defibrillator/Transcutaneous Pacemaker Analyzer Specifications 33 Automated Defibrillator Test ECG Waves Normal Sinus...
Impulse 6000D, 7000DP Users Manual 34 Lead V1...24 Lead V2...
Defibrillator/Transcutaneous Pacemaker Analyzer Specifications 35 Noise Immunity Wave ...
Returns for partial refund/credit: Every product returned for refund/credit must be accompanied by a Return Material Authorization (RMA) number, obta
Impulse 6000D, 7000DP Users Manual 36 Paroxysmal Atrial Tachycardia (PAT) Nodal Rhythm Supraventricular Tachycardia Premature Atrial PAC Nodal PNC PVC
Defibrillator/Transcutaneous Pacemaker Analyzer Specifications 37 Asystole Conduction 1° Block 2° Block Type I 2° Block Type II 3° Block Right
Impulse 6000D, 7000DP Users Manual 38 Transcutaneous Pacemaker Analyzer Specifications (Impulse 7000DP only) Test Load Selections Defibrillator Input
Defibrillator/Transcutaneous Pacemaker Analyzer Specifications 39 Pulse Rate Range...
Impulse 6000D, 7000DP Users Manual 40 Sensitivity Test Automatic Interactive Threshold Detection Compatible pacer rates...
41 Appendix A Impulse 6000D/7000DP Remote OperationAnsur Test Guide This appendix describes using the Ansur plug-in software to execute tests on the
Impulse 6000D, 7000DP Users Manual 42 fcz06.bmp Figure 30. Ansur Test Guide Window
Defibrillator/Transcutaneous Pacemaker Analyzer Defibrillator Tests 43 Defibrillator Tests The Impulse 6000D/7000DP Plug-in allows testomg of
Impulse 6000D, 7000DP Users Manual 44 fcz07.bmp Figure 31. Graph of Discharge Curve
Defibrillator/Transcutaneous Pacemaker Analyzer Defibrillator Tests 45 Charge Time Test The Charge Time test measures how long it takes to cha
Restrictions and Liabilities Information in this document is subject to change and does not represent a commitment by Fluke Biomedical. Changes made t
Impulse 6000D, 7000DP Users Manual 46 5. Discharge the defibrillator. Results of the test appear in the Test results pane. Pacemaker Tests Pacemaker
Defibrillator/Transcutaneous Pacemaker Analyzer Pacemaker Tests 47 vary in rate and making several timing measurements on how the pacer respon
Impulse 6000D, 7000DP Users Manual 48 Asynchronous Mode Test This test follows the same procedure as the Pacer Demand Mode but is run with the pacer i
Defibrillator/Transcutaneous Pacemaker Analyzer ECG Waveform Simulation Tests 49 the BPM is outside the limits specified in the test procedure
Impulse 6000D, 7000DP Users Manual 50 5. Click the Test passed checkbox or the Test failed checkbox to record the observed result of the test. ECG No
Defibrillator/Transcutaneous Pacemaker Analyzer Battery Performance Tests 51 5. When charging is complete, discharge the defibrillator. The f
Impulse 6000D, 7000DP Users Manual 52
53 Appendix B Impulse 6000D/7000DP Test TemplatesIntroduction This chapter introduces the template capabilities of the Impulse 6000D/7000DP Plug-In
Impulse 6000D, 7000DP Users Manual 54 fcz08.bmp Figure 32. Test Template with Selected Test Element
Defibrillator/Transcutaneous Pacemaker Analyzer Creating Test Templates 55 In the middle of the Test Template window are located the following
i Table of Contents Title Page Defibrillator Analyzer ...
Impulse 6000D, 7000DP Users Manual 56 3. Click the Apply when tab to assign report levels, standards, and service events to test elements. 4. Click
Defibrillator/Transcutaneous Pacemaker Analyzer Creating Test Templates 57 5. To define how Ansur calculates the limit values for certain mea
Impulse 6000D, 7000DP Users Manual 58 6. To add or delete limits, right click one of the rows of the Expected results page and select from the pop-up
Defibrillator/Transcutaneous Pacemaker Analyzer Creating Test Templates 59 7. Click the Custom setup tab to view and define the parameters us
Impulse 6000D, 7000DP Users Manual 60 Defibrillator tests also have available the Auto Advance option that is disabled by default. This option advance
Defibrillator/Transcutaneous Pacemaker Analyzer Using Defibrillator Test Elements 61 Table 5. Energy Measurement Test Measurements Measurement
Impulse 6000D, 7000DP Users Manual 62 Charge Time Test The Charge Time test measures how long a defibrillator takes to charge up to a specified energy
Defibrillator/Transcutaneous Pacemaker Analyzer Using Defibrillator Test Elements 63 Table 8. Charge Time Test Custom Parameters Parameter De
Impulse 6000D, 7000DP Users Manual 64 Table Synchronization Time Test Measurements (cont) Measurement Unit of Measure Description Pulse Width 50%
Defibrillator/Transcutaneous Pacemaker Analyzer Using Pacemaker Test Elements (Impulse 7000DP only) 65 Table 11. Pacer Parameter Test Measurem
7000DP, 6000D Users Manual ii Performing a Pacer Demand Test... 15 Perform
Impulse 6000D, 7000DP Users Manual 66 Pacer Refractory Test The Pacer Refractory test checks the ability of the pacemaker to interact with cardiac act
Defibrillator/Transcutaneous Pacemaker Analyzer Using Pacemaker Test Elements (Impulse 7000DP only) 67 Table 14. Pacer Refractory Test Custom
Impulse 6000D, 7000DP Users Manual 68 Table 16. Pacer Sensitivity Test Custom Parameters Parameter Description Input Jacks Specifies where the pacer
Defibrillator/Transcutaneous Pacemaker Analyzer Using Pacemaker Test Elements (Impulse 7000DP only) 69 Table 17. ECG Pacer Interactive Test Cu
Impulse 6000D, 7000DP Users Manual 70 Pacer Demand Mode Test The Pacer Demand Mode test is used to verify demand mode pacing over a range of BPM rates
Defibrillator/Transcutaneous Pacemaker Analyzer Using ECG Waveform Simulation Test Elements 71 Asynchronous Mode Test The Asynchronous Mode Te
Impulse 6000D, 7000DP Users Manual 72 Arrhythmia Wave Test This test is typically used to verify the shock advisory capability of a defibrillator in r
Defibrillator/Transcutaneous Pacemaker Analyzer Using ECG Waveform Simulation Test Elements 73 Performance Wave Simulation The Performance Wav
Impulse 6000D, 7000DP Users Manual 74 ECG R-Wave Test The ECG R-Wave test checks the ability of the ECG monitor to detect an R-Wave over a range of R-
Defibrillator/Transcutaneous Pacemaker Analyzer Using Battery Performance Test Elements 75 ECG Noise Immunity Test The ECG Noise Immunity test
Contents (continued) iii Appendix A - Impulse 6000D/7000DP Remote Operation ...41 Ansur Test
Impulse 6000D, 7000DP Users Manual 76 Table 24. Battery Capacity Test Measurements Measurement Unit of Measure Description Charge Time seconds Ho
Defibrillator/Transcutaneous Pacemaker Analyzer Using Battery Performance Test Elements 77 Defib Pulse Repetition Test The Defib Pulse Repetit
Impulse 6000D, 7000DP Users Manual 78