Summary of Contents

Page 1 - Nuclear Associates 76-025

Nuclear Associates 76-025 CDRH Dental Image Quality Test Tool Instruction Manual

Page 2

Radiation Management Services For additional information, please contact Radiation Management Services business of Cardinal Health at 440.248.9

Page 3 - Contents

(Blank Page)

Page 4 - (Blank Page)

i Contents Section 1: General Information ... 1-

Page 6 - 1.2 Setup Procedure

1-1Section 1 General Information 1.1 Introduction The Dental Image Quality Test Tool was designed to provide a means of testing the functionality of

Page 7 - Operation

1-21.2 Setup Procedure 1. Place the Dental Image Quality Test Tool cradle on a support. The phantom cradle should be placed so that it is level and

Page 8 - 2.4 HVL Exposures

2-1Section 2 Operation 2.1 Darkroom Fog Measurement The following procedure is used to sensitize a film for determining darkroom fog levels. You w

Page 9

2-2 2.2 Phantom Image Evaluation 1. Insert one packet of the facility's film into the film holder, which is a “U”-shaped area located in the

Page 10 - All rights reserved

2-32.5 kVp Determination by Copper Transmission Method 1. Use the same setup as you did for the HVL determination. 2. Select a technique that giv