medSim 300B
Operators Manual
4-28
Intrauterine Pressure (IUP) Simulation
The intrauterine pressure (IUP) waveform is output at "BP4/AUX".
The following drawing illustrates the medSim 300B simulated intrauterine pressure
waveform.
0 Seconds 90
10
90
mmHg
eic198.eps
Figure 4-23. medSim 300B Simulated Intrauterine Pressure Waveform
At the blood pressure "SENSITIVITY" switch located on the front panel, adjust the
medSim 300B to match the input sensitivity of the fetal monitor—either 5 µV/V/mmHg
or 40 µV/V/mmHg.
The frequency of the uterine contractions generated by the medSim 300B can be varied.
To vary the rate at which contractions occur:
1. Select <F1>-ECG from the FUNCTION SELECT MENU for the following two
pages of display:
ECG
BASE PERF ARRH >
<F1> <F2> <F3> <F4>
PVC
eic199.eps
ECG
< PACE INS FETAL
<F1> <F2> <F3> <F4>
eic200.eps