Nuclear Associates 07-492 Mammographic Digital kVp Meter Operators Manual March 2005 Manual No. 164001 Rev. 3 ©200
Nuclear Associates 07-492 Operators Manual 2-2 exposure. To improve accuracy, the user should either increase distance or decrease the mA, both of
Battery Installation and SetupInstalling/Replacing the Battery3 3-1Section 3 Battery Installation and Setup 3.1 Installing/Replacing the Battery U
Nuclear Associates 07-492 Operators Manual 3-2 Figure 3-1. Detector Positioning 3.4 Determining mAs Requirements The mAs mini
Battery Installation and SetupDetermining mAs Requirements3 3-3Typically, the largest apparent variation in mAs requirements will be due to a kVp e
Nuclear Associates 07-492 Operators Manual 3-4 If an oscilloscope is used to measure the voltage divider, an independent voltmeter should be us
Theory of OperationGeneral4 4-1Section 4 Theory of Operation 4.1 General The kVp measurement is computed basically from a measurement of the linea
Nuclear Associates 07-492 Operators Manual (Blank page)
Calibration and AdjustmentLo Battery Threshold Adjustment5 5-1Section 5 Calibration and Adjustment The Model 07-492 kVp Meter has been factory cal
Nuclear Associates 07-492 Operators Manual 5-2 Be sure the phase switch on the rear panel of the kVp meter is in the appropriate position. 3.
(Blank page)
Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.2564 120 Andrews R
Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.2564
iTable of Contents Section 1: General Information... 1-1 1.1 Produc
(Blank page)
General InformationProduct Description1 1-1Section 1 General Information Care must be taken to ensure radiation beam is off prior to placing th
Nuclear Associates 07-492 Operators Manual 1-2 Correction factors for different target/filter combinations are provided in Section 5, Calibrati
General InformationReceiving Inspection1 1-31.3 Receiving Inspection 1. Check the packaging and its contents for damage. If damage is evident, fil
Nuclear Associates 07-492 Operators Manual 1-4 terminate the warranty agreement between Fluke Biomedical, Radiation Management Services and the cu
Indicators and ControlsFront Panel2 2-1Section 2 Indicators and Controls 2.1 Front Panel See Figure 2-1 for front panel control and indicator loca