Nuclear Associates 07-492
Mammographic Digital kVp Meter
Operators Manual
March 2005
Manual No. 164001 Rev. 3
©2003, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A.
All product names are trademarks of their respective companies
Page view 0
1 2 3 4 5 6 ... 19 20

Summary of Contents

Page 1

Nuclear Associates 07-492 Mammographic Digital kVp Meter Operators Manual March 2005 Manual No. 164001 Rev. 3 ©200

Page 2

Nuclear Associates 07-492 Operators Manual 2-2 exposure. To improve accuracy, the user should either increase distance or decrease the mA, both of

Page 3 - Table of Contents

Battery Installation and SetupInstalling/Replacing the Battery3 3-1Section 3 Battery Installation and Setup 3.1 Installing/Replacing the Battery U

Page 4

Nuclear Associates 07-492 Operators Manual 3-2 Figure 3-1. Detector Positioning 3.4 Determining mAs Requirements The mAs mini

Page 5 - General Information

Battery Installation and SetupDetermining mAs Requirements3 3-3Typically, the largest apparent variation in mAs requirements will be due to a kVp e

Page 6 - 1.2 Specifications

Nuclear Associates 07-492 Operators Manual 3-4 If an oscilloscope is used to measure the voltage divider, an independent voltmeter should be us

Page 7 - 1.5 Routine Cleaning

Theory of OperationGeneral4 4-1Section 4 Theory of Operation 4.1 General The kVp measurement is computed basically from a measurement of the linea

Page 8

Nuclear Associates 07-492 Operators Manual (Blank page)

Page 9 - Indicators and Controls

Calibration and AdjustmentLo Battery Threshold Adjustment5 5-1Section 5 Calibration and Adjustment The Model 07-492 kVp Meter has been factory cal

Page 10 - 2.5 Rear Panel

Nuclear Associates 07-492 Operators Manual 5-2 Be sure the phase switch on the rear panel of the kVp meter is in the appropriate position. 3.

Page 12

Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.2564 120 Andrews R

Page 13 - 3.6 Voltage Divider Method

Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.2564

Page 14

iTable of Contents Section 1: General Information... 1-1 1.1 Produc

Page 16 - (Blank page)

General InformationProduct Description1 1-1Section 1 General Information Care must be taken to ensure radiation beam is off prior to placing th

Page 17 - Calibration and Adjustment

Nuclear Associates 07-492 Operators Manual 1-2 Correction factors for different target/filter combinations are provided in Section 5, Calibrati

Page 18

General InformationReceiving Inspection1 1-31.3 Receiving Inspection 1. Check the packaging and its contents for damage. If damage is evident, fil

Page 19

Nuclear Associates 07-492 Operators Manual 1-4 terminate the warranty agreement between Fluke Biomedical, Radiation Management Services and the cu

Page 20

Indicators and ControlsFront Panel2 2-1Section 2 Indicators and Controls 2.1 Front Panel See Figure 2-1 for front panel control and indicator loca