PN 2572314 March 2006, Rev. 1, 4/15 © 2006-2015 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All pr
DPM1B Operators Manual vi
1 Introduction This document is the Operators manual for the DPM1B Pneumatic Transducer Tester (hereafter “the Tester”). It contains general informat
DPM1B Operators Manual 2 • The battery door must be closed and locked before you operate the Product. • Replace the battery when the low battery ind
Pneumatic Transducer Tester Introduction 3 Symbols Table 1 is a list of symbols found in this manual or on the Tester. Table 1. Symbols Symbol Desc
DPM1B Operators Manual 4 The Tester has the following features: • Easy-to-read liquid crystal display (LCD) • Capable of measuring and generating po
Pneumatic Transducer Tester Specifications 5 the transducer. Unfortunately, manometers can add significant errors themselves, because of the diffic
DPM1B Operators Manual 6 Electromagnetic Compatibility (EMC) International ... IEC 61326-1: Porta
Pneumatic Transducer Tester Accessories 7 ON/OFF SwitchThumbwheelLCDLuer LockConnectionSiteZERO Knob ehf1.eps Figure 1. DPM1B Front Panel Back Pane
DPM1B Operators Manual 8 Operation Overview When operating according to the instructions given in this section, the Tester produces and displays a pre
Pneumatic Transducer Tester Operation Overview 9 PneumaticTransducerTesterBP InputFilterMonitor21 ehf2.eps Figure 2. DPM1B to Monitor Setup 3. Whe
Warranty and Product Support Fluke Biomedical warrants this instrument against defects in materials and workmanship for one full year from the date
DPM1B Operators Manual 10 Note Allow approximately 3 minutes warm-up time before taking readings; this allows the electronics to stabilize in the Test
Pneumatic Transducer Tester Testing Transducer Linearity 11 To interpret the results, proceed to the following three sections: • Testing Transduc
DPM1B Operators Manual 12 Transducer Sensitivity Test Only one measurement is required to determine whether or not a transducer under test meets the m
Pneumatic Transducer Tester Testing Other Devices 13 Testing Other Devices A variety of pressure-measuring and pressure-generating devices can be
DPM1B Operators Manual 14 Maintenance Warning For safe operation and maintenance of the product and to prevent possible electrical shock, fire, or
Notices All Rights Reserved Copyright 2006-2015, Fluke Biomedical. No part of this publication may be reproduced, transmitted, transcribed, stored
Use the original carton and packaging material for shipment. If they are not available, we recommend the following guide for repackaging: Use a d
i Table of Contents Title Page Introduction ... 1 Safety ...
DPM1B Operators Manual ii
iii List of Tables Table Title Page 1. Symbols ... 3 2. Accessories ...
DPM1B Operators Manual iv
v List of Figures Figure Title Page 1. DPM1B Front Panel ............... 7 2. DPM1B to Monitor Setup