Victoreen® 1060AM Digital Smart Detector Area Monitor Operators Manual March 2005 Manual No. 1060AM-1 Rev. 2 ©2004,

Page 2 - 440.498.2564

Victoreen 1060AM Operators Manual 1-6 Resolving time: 1 microsecond Plateau Slope: 2% per 100 V Tube Life expectancy: >1010 counts Neutron Sen

Page 3 - Table of Contents

Victoreen 1060AM Operators Manual 1-6 The Model 1060-xxx scintillation detector does not contain an integral preamplifier, the Model 943-36 does.

Page 4

Victoreen 1060AM Operators Manual (Blank Page)

Page 5 - Introduction

Installation/ReplacementInstallation2 2-1CAUTIONCAUTIONCAUTIONSection 2 Installation/Replacement 2.1 Installation The device power must be turned OF

Page 6 - 1.2 Specifications

Victoreen 1060AM Operators Manual 2-2 J3 RS-485 Connector (4-pin Female) J4 RS-485 Connector (4-pin Female) 1 A 2 B 3 GND 4 Termination (120Ω) RS

Page 7 - 1.2.2 Analog Specifications

Installation/ReplacementInstallation2 2-313. Some commercial RS-485 to RS-232 converters require the external communications device (1060) to be conn

Page 8

Victoreen 1060AM Operators Manual 2-4 2.4.1 Input Channel Selection The Select Input Channel command (SIC) allows one of the four input processing

Page 9 - 1.3 Detector Types

OperationCommunication Protocol3 3-1Section 3 Operation 3.1 Communication Protocol Communication Port Command Entry Syntax RS-232 Communication Por

Page 10 - Specifications

Victoreen 1060AM Operators Manual 3-2 3.2.1 COMMAND SET The COMMAND SET is comprised of STANDARD, SETUP, TEST, and RADIATION commands. STANDARD comm

Page 11 - 1.5 Storage

OperationTerminal Operation3 3-3NOTERRS r Read mR/Hr (latest second mR/Hr value) RSH r Read Sieverts (averaged over 60 minutes) RSM r Read S

Page 12

Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.256

Page 13 - Installation/Replacement

Victoreen 1060AM Operators Manual 3-4 7. Open 8. ON = Terminal vs Computer communication format (Reserved) 3.5 Data Output The standard communic

Page 14

Theory of OperationGeneral4 4-1Section 4 Theory of Operation 4.1 General 4.1.1 Detector Input The Model 1060 has four separate input terminal points,

Page 15 - 2.3 Enclosures

Victoreen 1060AM Operators Manual 4-2 4.1.7 Digital inputs There are 16 TTL level digital inputs with pull up resistors. They may be used to indicat

Page 16 - Figure 2-3. Block Diagram

Calibration, Maintenance and TroubleshootingCalibration5 5-1CAUTIONCAUTIONCAUTIONSection 5 Calibration, Maintenance and Troubleshooting 5.1 Calibrat

Page 17 - Operation

Victoreen 1060AM Operators Manual 5-2 CAUTIONCAUTIONCAUTION5.3 Corrective Maintenance 5.3.1 Board Replacement To be provided later. 5.3.2 Detector

Page 18 - 3.2.1 COMMAND SET

Calibration, Maintenance and TroubleshootingReplacement Parts5 5-35.5 Replacement parts Part Number Description 1060-100-10 Model 1060 Printed Ci

Page 19 - 3.4 DIP Switch Definitions

Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.256

Page 20 - 3.7 Counts

iTable of Contents Section 1: Introduction... 1-1 1.1 P

Page 21 - Theory of Operation

ii3.7.4 Counts in mR/Hr... 3-4 3.7.5 Counts in Sieverts ...

Page 22 - 4.3 Circuit Description

IntroductionProduct Description1 1-1Section 1 Introduction 1.1 Product Description The Model 1060 is a smart radiation detector. It is a printed cir

Page 23 - Exposure

Victoreen 1060AM Operators Manual 1-2 NEMA 4X Enclosure Model No. Description 1060AM-NM-ER Environmental range - 1 μR to 1000 (μR/hr (SI units) 1

Page 24 - 5.4 Troubleshooting

IntroductionSpecifications1 1-3Digital Inputs 16 digital inputs with internal pull-up resistors 1.2.2 Analog Specifications High Voltage Regulated

Page 25 - 5.5 Replacement parts

Victoreen 1060AM Operators Manual 1-4 System Operation As part of a larger system, the 1060 will communicate via serial link to a host computer. I

Page 26

IntroductionSpecifications1 1-5• Provide digital input and analog inputs signals, such as for pressure and flow 10. Responds to data entry swit