Summary of Contents

Page 1 - Users Manual

PN FBC-0039 April 2013, Rev. 2, 10/14 © 2013-2014 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All p

Page 2

IDA-5 Users Manual vi

Page 3 - Notices

1 Introduction The Fluke Biomedical IDA-5 Infusion Device Analyzer (the Product) is a precision instrument that examines the performance of medical i

Page 4 - Manufacturing Location

IDA-5 Users Manual 2 Safety Information A Warning identifies hazardous conditions and actions that could cause bodily harm or death. A Caution identif

Page 5 - Table of Contents

Infusion Device Analyzer Safety Information 3 • Avoid possible contamination of reusable components due to backflow conditions. Some older style

Page 6

IDA-5 Users Manual 4  Caution To prevent possible damage to the product or to equipment under test: • Only qualified service personnel should servic

Page 7 - List of Tables

Infusion Device Analyzer Instrument Familiarization 5 Instrument Familiarization Tables 2 and 3 tell you about the controls and connections on the

Page 8

IDA-5 Users Manual 6 Table 3. Rear-Panel Controls and Connections 2134567 gir02.epsItem Description  Handle  Power switch  Power inlet  Equipotent

Page 9 - List of Figures

Infusion Device Analyzer Product Connections 7 Product Connections The Product connects to infusion devices through the front-panel inlet ports. F

Page 10

IDA-5 Users Manual 8  Caution Do not use delivery set or components that have been used for prior testing for patient infusion. Note Before you use t

Page 11 - Unpack the Product

Infusion Device Analyzer Product Operation 9 Keyboard It is recommended that a small footprint USB keyboard be used with the Product. The keyboard

Page 12 - Safety Information

Warranty and Product Support Fluke Biomedical warrants this instrument against defects in materials and workmanship for one year from the date of orig

Page 13 - Infusion Device Analyzer

IDA-5 Users Manual 10 Preferences To set your personal preferences, use the Utilities menu. To access this menu, select ESC while the Status All Chann

Page 14 - Symbols

Infusion Device Analyzer Product Operation 11 How to Test Infusion Devices When you test an infusion device, it is recommended that you record the

Page 15 - Instrument Familiarization

IDA-5 Users Manual 12 Select Device Information to record data for the test. The Device Information screen discussed in Table 5 is shown. Initially, o

Page 16

Infusion Device Analyzer Product Operation 13 When the data fields are filled in, the menu in Table 6 lets you choose what to do next. Table 6. D

Page 17 - Product Connections

IDA-5 Users Manual 14 Flow Tests To do a flow test, select Flow on the menu and then ENTER to open the Channel Flow screen shown in Figure 5. The scre

Page 18 - Connect Accessories

Infusion Device Analyzer Flow Tests 15 Table 8. Flow Screen Menu Controls Channel 1 Flow Average Flow 0.00 ml/h Volume 0.00 Ml Elapsed Time 00:

Page 19 - Product Operation

IDA-5 Users Manual 16 Select Graph on the menu and then push ENTER to show a graph of the test progress as shown in Figure 6 prime mode. Push End and

Page 20

Infusion Device Analyzer Flow Tests 17 Table 11. Flow Screen in End Mode Channel 1 Flow Average Flow 59.70 ml/h Volume 20.00 ml Elapsed Time 00:2

Page 21

IDA-5 Users Manual 18 Occlusion Tests When Occl is selected from the Device Information screen, the Occlusion Start Mode screen in Table 12 is shown.

Page 22 - Device Information Channel 1

Infusion Device Analyzer Occlusion Tests 19 When you select Start, the Channel Occlusion screen in Table 13 shows measurement data as the test con

Page 23

Notices All Rights Reserved  Copyright 2014, Fluke Biomedical. No part of this publication may be reproduced, transmitted, transcribed, stored in a r

Page 24 - Flow Tests

IDA-5 Users Manual 20 Table 14. Occlusion Screen in End Mode Channel 1 Occlusion Peak Pressure 8.04 psi Time of Peak 00: 07: 23 Total Test Time 00:

Page 25 - Volume 0.00

Infusion Device Analyzer How to Test PCA Pumps 21 How to Test PCA Pumps When PCA is selected from the Device Information screen, the PCA/Dual Flow

Page 26 - Press ESC for Status Screen

IDA-5 Users Manual 22 When ENTER is pushed in the Loading Dose field, the PCA screen changes to the Prime Mode screen shown in Figure 8. Refer to the

Page 27 - Volume 20.00

Infusion Device Analyzer Dual Flow Pump Test 23 Channel 1 PCA/Dual Flow Graph 80 Ave ml/hBolus Count 3 Basal Avg 0 Bolus Avg 65.00 Tota

Page 28 - Occlusion Tests

IDA-5 Users Manual 24 Define a Template To define a template: 1. From the Status All Channels screen, select ESC for Utilities. 2. Select Edit Templ

Page 29 - Time 00:09:10

Infusion Device Analyzer Templates 25 5. Repeat for steps 2 and 3. Use the values shown in Table 17. 6. At step 4, keep the Type field empty and

Page 30

IDA-5 Users Manual 26 Utilities To open the Utilities menu screen shown in Figure 10, select ESC from the Status All Channels screen. The subsequent s

Page 31 - How to Test PCA Pumps

Infusion Device Analyzer Utilities 27 At the bottom of the screen, there are context-sensitive prompts. Use the up and down arrows to highlight or

Page 32

IDA-5 Users Manual 28 The Product can automatically do tests in a specific manner when specified conditions are met. Test-stop preferences specifies t

Page 33 - Templates

Infusion Device Analyzer Troubleshooting 29 Troubleshooting Tables 19 and 20 show you Bubble and Air Lock errors and their solutions. Table 19. Bu

Page 34 - Save Delete Step

Certification This instrument was thoroughly tested and inspected. It was found to meet Fluke Biomedical’s manufacturing specifications when it was sh

Page 35

IDA-5 Users Manual 30 Product Maintenance The subsequent sections tell you how to maintain the Product. Clean the Product  Warning To prevent possib

Page 36 - Utilities

Infusion Device Analyzer Product Maintenance 31 Replaceable Parts Table 21 lists the replaceable parts for the Product. Table 21. Replaceable Part

Page 37 - Default Test Preferences

IDA-5 Users Manual 32 Test Fluid The Product is intended to operate with de-ionized water with added detergent. Fluids intended for use on patients, s

Page 38

Infusion Device Analyzer General Specifications 33 General Specifications Operating Voltage Range ... 100 V ac t

Page 39 - Troubleshooting

IDA-5 Users Manual 34

Page 40 - Product Maintenance

i Table of Contents Title Page Introduction ... 1 Inten

Page 41 - Replaceable Parts

IDA-5 Users Manual ii Clean the Product ... 30 Outside ...

Page 42

iii List of Tables Table Title Page 1. Symbols ...

Page 44

v List of Figures Figure Title Page 1. Infusion Device Connections to the Product ............ 7 2. Drain Conn