PN FBC-0039 April 2013, Rev. 2, 10/14 © 2013-2014 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All p
IDA-5 Users Manual vi
1 Introduction The Fluke Biomedical IDA-5 Infusion Device Analyzer (the Product) is a precision instrument that examines the performance of medical i
IDA-5 Users Manual 2 Safety Information A Warning identifies hazardous conditions and actions that could cause bodily harm or death. A Caution identif
Infusion Device Analyzer Safety Information 3 • Avoid possible contamination of reusable components due to backflow conditions. Some older style
IDA-5 Users Manual 4 Caution To prevent possible damage to the product or to equipment under test: • Only qualified service personnel should servic
Infusion Device Analyzer Instrument Familiarization 5 Instrument Familiarization Tables 2 and 3 tell you about the controls and connections on the
IDA-5 Users Manual 6 Table 3. Rear-Panel Controls and Connections 2134567 gir02.epsItem Description Handle Power switch Power inlet Equipotent
Infusion Device Analyzer Product Connections 7 Product Connections The Product connects to infusion devices through the front-panel inlet ports. F
IDA-5 Users Manual 8 Caution Do not use delivery set or components that have been used for prior testing for patient infusion. Note Before you use t
Infusion Device Analyzer Product Operation 9 Keyboard It is recommended that a small footprint USB keyboard be used with the Product. The keyboard
Warranty and Product Support Fluke Biomedical warrants this instrument against defects in materials and workmanship for one year from the date of orig
IDA-5 Users Manual 10 Preferences To set your personal preferences, use the Utilities menu. To access this menu, select ESC while the Status All Chann
Infusion Device Analyzer Product Operation 11 How to Test Infusion Devices When you test an infusion device, it is recommended that you record the
IDA-5 Users Manual 12 Select Device Information to record data for the test. The Device Information screen discussed in Table 5 is shown. Initially, o
Infusion Device Analyzer Product Operation 13 When the data fields are filled in, the menu in Table 6 lets you choose what to do next. Table 6. D
IDA-5 Users Manual 14 Flow Tests To do a flow test, select Flow on the menu and then ENTER to open the Channel Flow screen shown in Figure 5. The scre
Infusion Device Analyzer Flow Tests 15 Table 8. Flow Screen Menu Controls Channel 1 Flow Average Flow 0.00 ml/h Volume 0.00 Ml Elapsed Time 00:
IDA-5 Users Manual 16 Select Graph on the menu and then push ENTER to show a graph of the test progress as shown in Figure 6 prime mode. Push End and
Infusion Device Analyzer Flow Tests 17 Table 11. Flow Screen in End Mode Channel 1 Flow Average Flow 59.70 ml/h Volume 20.00 ml Elapsed Time 00:2
IDA-5 Users Manual 18 Occlusion Tests When Occl is selected from the Device Information screen, the Occlusion Start Mode screen in Table 12 is shown.
Infusion Device Analyzer Occlusion Tests 19 When you select Start, the Channel Occlusion screen in Table 13 shows measurement data as the test con
Notices All Rights Reserved Copyright 2014, Fluke Biomedical. No part of this publication may be reproduced, transmitted, transcribed, stored in a r
IDA-5 Users Manual 20 Table 14. Occlusion Screen in End Mode Channel 1 Occlusion Peak Pressure 8.04 psi Time of Peak 00: 07: 23 Total Test Time 00:
Infusion Device Analyzer How to Test PCA Pumps 21 How to Test PCA Pumps When PCA is selected from the Device Information screen, the PCA/Dual Flow
IDA-5 Users Manual 22 When ENTER is pushed in the Loading Dose field, the PCA screen changes to the Prime Mode screen shown in Figure 8. Refer to the
Infusion Device Analyzer Dual Flow Pump Test 23 Channel 1 PCA/Dual Flow Graph 80 Ave ml/hBolus Count 3 Basal Avg 0 Bolus Avg 65.00 Tota
IDA-5 Users Manual 24 Define a Template To define a template: 1. From the Status All Channels screen, select ESC for Utilities. 2. Select Edit Templ
Infusion Device Analyzer Templates 25 5. Repeat for steps 2 and 3. Use the values shown in Table 17. 6. At step 4, keep the Type field empty and
IDA-5 Users Manual 26 Utilities To open the Utilities menu screen shown in Figure 10, select ESC from the Status All Channels screen. The subsequent s
Infusion Device Analyzer Utilities 27 At the bottom of the screen, there are context-sensitive prompts. Use the up and down arrows to highlight or
IDA-5 Users Manual 28 The Product can automatically do tests in a specific manner when specified conditions are met. Test-stop preferences specifies t
Infusion Device Analyzer Troubleshooting 29 Troubleshooting Tables 19 and 20 show you Bubble and Air Lock errors and their solutions. Table 19. Bu
Certification This instrument was thoroughly tested and inspected. It was found to meet Fluke Biomedical’s manufacturing specifications when it was sh
IDA-5 Users Manual 30 Product Maintenance The subsequent sections tell you how to maintain the Product. Clean the Product Warning To prevent possib
Infusion Device Analyzer Product Maintenance 31 Replaceable Parts Table 21 lists the replaceable parts for the Product. Table 21. Replaceable Part
IDA-5 Users Manual 32 Test Fluid The Product is intended to operate with de-ionized water with added detergent. Fluids intended for use on patients, s
Infusion Device Analyzer General Specifications 33 General Specifications Operating Voltage Range ... 100 V ac t
IDA-5 Users Manual 34
i Table of Contents Title Page Introduction ... 1 Inten
IDA-5 Users Manual ii Clean the Product ... 30 Outside ...
iii List of Tables Table Title Page 1. Symbols ...
IDA-5 Users Manual iv
v List of Figures Figure Title Page 1. Infusion Device Connections to the Product ............ 7 2. Drain Conn
Comments to this Manuals