Nuclear Associates 76-903 MRI Multi-Purpose Phantom Users Manual March 2005 Manual No. 76-903-1 Rev. 2 ©2004, 200
Nuclear Associates 76-903 Operators Manual 3-2 Figure 3-2 shows the cross section of the concentric conical part of the phantom. The aspect ratio is
Theory of OperationFlood Section3 3-3 Figure 3-4. Diameter, 9", Depth, 1.25" 3.3.1 RF Uniformity The flood section is used to mea
Nuclear Associates 76-903 Operators Manual 3-4 3. Star Pattern: Modulation transfer function (MTF), horizontal and vertical spatial resolution, qua
Nuclear Associates 76-903 Operators Manual 3-6 Figure 3-7. 3.4.4 Slice Continuity (Gap) The slice contiguity can be evaluated by quick v
Nuclear Associates 76-903 Operators Manual 3-6 Figure 3-9. 0.5 mm x 0.5 mm2, 0.75 mm x 0.75 mm2, 1.0 mm x 1.0mm2, and 2.0 mm x 2.0mm2 square holed
Theory of OperationInsert Section3 3-7The star pattern for MTF evaluation is made up of 3° wedges placed on a 60° fan angle. Each wedge is 3” in the
Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.256
Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.256
iTable of Contents Section 1: Introduction... 1-1 1.1
(Blank Page)
IntroductionIntroduction1 1-1Section 1 Introduction 1.1 Introduction This phantom is designed for comprehensive evaluation of critical imaging param
Nuclear Associates 76-903 Operators Manual 1-2 1.4 NMR Solution Any NMR signal-producing solution can be used to fill the phantom. It is desirable
SpecificationsPhantom Design2 2-1Section 2 Specifications 2.1 Phantom Design All of the parts in the phantom are made with plastic and glass, free o
Nuclear Associates 76-903 Operators Manual 2-2 The insert section has six inserts, one 120° star pattern wedge assembly, two folding step wedges, sp
Theory of OperationOmni Directional Application3 3-1Section 3 Theory Of Operation 3.1 Omni Directional Application Since MR images are obtained not