ESA620
Users Manual
4
The Analyzer must be properly earthed.
Only use a supply socket that has a
protective earth contact. If there is any
doubt as to the effectiveness of the
supply socket earth, do not connect the
Analyzer. Do not use a two-conductor
adapter or extension cord; this will break
the protective ground connection.
Use extreme caution when working with
voltages above 30 volts.
Use the proper terminals, functions and
ranges for the test being performed.
Do not touch metal parts of the device
under test (DUT) during analysis. The
DUT should be considered an electrical
shock hazard when connected to the
Analyzer as some tests involve high
voltages, high currents, and/or the
removal of DUT earth bond.
Intended Use
The Analyzer is intended for use by trained service
technicians to perform periodic inspections on a wide
range of medical equipment. The testing procedures are
menu-driven, and simple to operate.
The Product is an electronic signal source and
measurement device for verifying the electrical safety of
medical devices. The Product also provides ECG
simulation and performance waveforms to verify patient
monitors are performing within their operating
specifications.
The Product provides the following function categories:
ECG Functions
ECG-Performance Testing
The intended user is a trained biomedical equipment
technician who performs periodic preventative
maintenance checks on patient monitors in service. Users
can be associated with hospitals, clinics, original
equipment manufacturers and independent service
companies that repair and service medical equipment.
The end user is an individual, trained in medical
instrumentation technology. This Product is intended to
be used in the laboratory environment, outside of the
patient care area, and is not intended for use on patients,
or to test devices while connected to patients. This
Product is not intended to be used to calibrate medical
equipment. It is intended for over the counter use.