Performing an Alternative Equipment Leakage
Test
Note
The alternative equipment leakage test is
available when the EN62353 & VDE 751
standard is selected.
During the Alternative Equipment Leakage test, the
voltage source is applied between short-circuited
equipment outlet mains live, neutral, and equipment outlet
earth, the exposed conductive surface on the housing,
and all applied parts short-circuited together. Equipment
is separated from mains during the test. The current
which flows over the insulation of the DUT is measured.
This test is not applicable for equipment with internal
electrical power source. The switches in mains part shall
be closed during measurement.
To perform an alternative equipment leakage test:
1. Press .
The alternative equipment test is the default test and
should already be selected.
2. Press to apply the voltage and read the current
in the display.
Figure 25 shows the electrical connections between the
Analyzer and the DUT during a Alternative Equipment
Leakage Test.
The following outlet conditions apply when performing this
test:
Closed Earth
Open Earth
Performing an Alternative Applied Part Leakage
Test
Note
The Alternative applied part leakage test is
available when the EN62353 & VDE 751
standard is selected.
During the Alternative Applied Part Leakage test, the test
voltage is applied between short-circuited applied parts of
a single function and the short-circuited equipment outlet
mains live, neutral, equipment outlet earth, and exposed
conductive surface on the housing. This test should only
be done for equipment with F-Type applied parts. For
equipment with multiple applied parts, test each group of
applied parts of a single function in turn with all others
floating during the test
. All applied parts can be
connected to the Analyzer’s applied parts jacks and the
lead selection will float those not selected.