Fluke-biomedical medTester 5000C User Manual

Browse online or download User Manual for Equipment Fluke-biomedical medTester 5000C. Fluke Biomedical medTester 5000C User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 292
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews

Summary of Contents

Page 1 - Operators Manual

medTester 5000C Biomedical Tester Operators Manual PN 2243153 February 2006  2006 Fluke Corporation, All rights reserved. Printed in USA All product

Page 2 - Warranty and Product Support

Test Sequence... 9-7 Posttest Prompts...

Page 3 - Notices

medTester 5000C Operators Manual 6-8 Customizing Your Autosequence Prompts In an autosequence, the medTester 5000C prompts you for information that is

Page 4 - Return Procedure

Biomedical Test System Customize Your medTester 6 6-9 medTester Autosequence Prompts Prompt # Prompt Name/ Renameable? (Yes or No) Description

Page 5

medTester 5000C Operators Manual 6-10 Select OFF, or ON. Then press NEXT to move to prompt two. If this is the only prompt or the last prompt that you

Page 6

Biomedical Test System Customize Your medTester 6 6-11 Pausing Between tests in a safety autosequence, you can insert a pause for three, six, o

Page 7

medTester 5000C Operators Manual 6-12 Any or all of these fields can be turned on (to appear on the test tag) or turned off on the printed test tag. W

Page 8

Biomedical Test System Customize Your medTester 6 6-13 RECSUMF1CUSTOMF3 The medTester displays the number of the first field to appear on each

Page 9

medTester 5000C Operators Manual 6-14 Follow directions, such as cycling power after reprogramming the scanner. The scanner remains programmed after c

Page 10

Biomedical Test System Customize Your medTester 6 6-15 Resetting Autosequences to Default Settings You can always return medTester 5000C autoseq

Page 11

medTester 5000C Operators Manual 6-16

Page 12

7-1 Chapter 7 Performance Waves In this chapter, you learn how to output performance waveforms to test patient monitors using the medTester 5000C. O

Page 13 - General Information

Biomedical Test System Contents Local Input Mode ...12-3 Local Output for Re

Page 14 - Introduction

medTester 5000C Operators Manual 7-2 Waveform Groups You can generate any of the waveforms listed in the five groups in the table below for the testin

Page 15 - How to Use This Manual

Biomedical Test System Performance Waves 7 7-3 WAVEFORM GROUPS medTester 5000C Display Wave Menu Name Description Remote Command PAIR A Pair o

Page 16

medTester 5000C Operators Manual 7-4 Running Waveforms from The WAVES Menu To turn on any of the waveforms, follow these instructions: 1. Attach the

Page 17 - Features

Biomedical Test System Performance Waves 7 7-5 Conducting Waveform Tests in Safety Autosequences You can execute performance wave tests at the e

Page 18

medTester 5000C Operators Manual 7-6

Page 19

8-1 Chapter 8 Memory This chapter acquaints you with the medTester 5000C memory module options you may have installed on your medTester. Introductio

Page 20

medTester 5000C Operators Manual 8-2 Note All medTester memory functions begin on MENU2 as described above. Viewing A Single Record You can view reco

Page 21

Biomedical Test System Memory 8 8-3 Print A Range of Records 1. Enter the menu command: RANGEF2 2. The display prompts you to PRINT FROM RECO

Page 22

medTester 5000C Operators Manual 8-4 Next access the type of records you want to print. From menu one, you can access these types: • SAFETYF1 Prints

Page 23 - Optional Accessories

Biomedical Test System Memory 8 8-5 Printing A Summary of Records You can print a summary of all records stored from the menu command: SUMMARYF5

Page 24

For the Oxitest PLUS: ... 15-13 For CardioSat 100: ...

Page 25 - AAMI Load

medTester 5000C Operators Manual 8-6 Deleting Records If you want, you can delete a single record with this command: DELETEF3 The medTester asks you t

Page 26 - CSA Label

Biomedical Test System Memory 8 8-7 1. Free Memory—In bytes. 2. Total Memory and Checklists—Amount of total memory and the amount used by check

Page 27

medTester 5000C Operators Manual 8-8 That indicates the number of the first record which contains the string. You can view or print the data in the fo

Page 28 - General Specifications

Biomedical Test System Memory 8 8-9 Keyboard Shortcut Commands In addition to accessing memory functions through menu commands, you can also ent

Page 29 - Accessories

medTester 5000C Operators Manual 8-10

Page 30 - Optional (continued)

9-1 Chapter 9 Defibrillator Module In this chapter you learn about using the medTester 5000C to control defibrillator autosequence tests with Fluke

Page 31

medTester 5000C Operators Manual 9-2 Note Consult your Impulse 3000 or Impulse 4000 operators manual for proper operation of your Impulse model. To di

Page 32

Biomedical Test System Defibrillator Module 9 9-3 Defibrillator Autosequence Names You can begin any of the autosequences listed below from the

Page 33 - Installation

medTester 5000C Operators Manual 9-4 DEFIBRILLATOR AUTOSEQUENCES Autosequence Name Keyboard Shortcut Command Menu Commands from DEFIBF2AUTOF1 M/D2J

Page 34 - Power-Up Sequence

Biomedical Test System Defibrillator Module 9 9-5 Running Defibrillator Autosequence Tests To begin running defibrillator autosequence tests, co

Page 35 - Date and Time Setup

1-1 Chapter 1 General Information In this chapter you will learn how to use this manual, where to get help, and about medTester 5000C Features and

Page 36 - Audio Transducer

medTester 5000C Operators Manual 9-6 Pretest Device Prompts There are eight prompts for information that can appear in the autosequence test record. A

Page 37 - Enabling Modules

Biomedical Test System Defibrillator Module 9 9-7 Note See Chapter 5, Autosequences, for a complete explanation of working with prompts. You can

Page 38

medTester 5000C Operators Manual 9-8 Sternum Paddle. The sternum paddle test is similar to the APEX paddle test. Two (2) leakage measurements are take

Page 39

Biomedical Test System Defibrillator Module 9 9-9 Out-of-Limits Reading When a reading is out of limits, a # sign appears next to the reading wi

Page 40

medTester 5000C Operators Manual 9-10 MAX Energy Test. This test measures the maximum length of time the EUT requires to charge to its maximum energy

Page 41 - Instrument Familiarity

Biomedical Test System Defibrillator Module 9 9-11 The table below supplies specific information about the tests for each autosequence. Autosequ

Page 42 - MED TESTER 5000

medTester 5000C Operators Manual 9-12 Posttest Prompts When all tests are completed, the medTester 5000C can prompt you for: 1. COMMENTS 2. NEXT TES

Page 43

Biomedical Test System Defibrillator Module 9 9-13 Begin by accessing the custom, autosequence, defib menu as described earlier. This time, sele

Page 44

medTester 5000C Operators Manual 9-14 Reinitialize Factory Default Settings Any customized autosequences can be returned to the factory default settin

Page 45 - 00/00/00 MENU 2 00:00:00

10-1 Chapter 10 IV Pump Module In this chapter you learn about using the medTester 5000C to control IV pump autosequence tests with a Fluke Biomedic

Page 46

medTester 5000C Operators Manual 1-2 Introduction This manual is written for the biomedical technician or clinical engineer responsible for testing ho

Page 47 - Manual Tests

medTester 5000C Operators Manual 10-2 Use the CUSTOM / AUTOSEQ / IVPUMP / TYPE menu path to configure the medTester for a specific IV pump analyzer.

Page 48

Biomedical Test System IV Pump Module10 10-3 Infutest 2000 Series D The Infutest can run three types tests: single rate tests, dual rate tests,

Page 49 - • NEUTRAL—open or closed

medTester 5000C Operators Manual 10-4 Note In order to take advantage of IV pump autosequences, your medTester 5000C requires the IV Pump Module, RS-

Page 50 - • Lead to lead

Biomedical Test System IV Pump Module10 10-5 The above display format and testing protocol is formatted for the Infutest 2000 Series D IV Pump A

Page 51 - TEST RECEPTACLE

medTester 5000C Operators Manual 10-6 IV Pump Autosequences for IPT-1 Autosequence Name (Type of Test) Keyboard Shortcut Command Menu Commands from

Page 52

Biomedical Test System IV Pump Module10 10-7 IV Pump Autosequences for IPT-MC Autosequence Name (Type of Test) Keyboard Shortcut Command Menu C

Page 53

medTester 5000C Operators Manual 10-8 Running IV Pump Autosequences To begin running IV pump autosequence tests, connect your medTester 5000C to your

Page 54

Biomedical Test System IV Pump Module10 10-9 1. OP CODE—This is the operator code for the person doing the testing. This code can consist of a m

Page 55 - TEST RECEPTACLE at the

medTester 5000C Operators Manual 10-10 IV PUMP TESTS After you respond to prompts, the autosequence tests begin. For each step in the autosequence, th

Page 56

Biomedical Test System IV Pump Module10 10-11 Dual Rate Test • The dual rate test measures the delivery of two sequential fluid volumes pumped

Page 57 - RESET (F1) key

Biomedical Test System General Information 1 1-3 The medTester 5000C—reliable safety and performance testing. Capable of managing your entire eq

Page 58

medTester 5000C Operators Manual 10-12 • The medTester displays test results on 3 screens. Press F5 NEXT to go to the next screen. Finally, you can r

Page 59

Biomedical Test System IV Pump Module10 10-13 IPT-1 • With the IPT-1, you can perform 2 types of flow tests: steady flow and high accuracy. Use

Page 60 - Case External Leakage

medTester 5000C Operators Manual 10-14 • For each test step, the medTester prompts you with the test type and conditions. Press F5 NEXT to continue.

Page 61 - Case Internal Leakage

Biomedical Test System IV Pump Module10 10-15 • Wait for the IDA 4 Plus to display PRIME on channel 1. Using a priming syringe, prime IV line i

Page 62 - External Leakage Current

medTester 5000C Operators Manual 10-16 • In a dual rate test, the medTester starts testing the secondary rate, indicated by “SEC” in the upper right

Page 63 - Manual Tests

Biomedical Test System IV Pump Module10 10-17 • The IDA 4 Plus will measure delivered bolus volumes ≥ 0.5 ml with no basal (or background) flui

Page 64 - ECG Interlead Leakage

medTester 5000C Operators Manual 10-18 Pressure Test (Infutest, IPT-1, IPT-MC and IDA 4 Plus) • After completing the flow test, the autosequence cont

Page 65

Biomedical Test System IV Pump Module10 10-19 INFUTEST 2000 Series D AUTOSEQUENCE TESTS Nominal Flow Rate (ml per hour) Test Type High Limit (

Page 66 - Equipment Load Current

medTester 5000C Operators Manual 10-20 IPT-1 AUTOSEQUENCE TESTS Nominal Flow Rate (ml per hour) Test Type High Limit (ml per hour) Low Limit (ml per

Page 67 - Power Cord Resistance Test

Biomedical Test System IV Pump Module10 10-21 IPT-MC AUTOSEQUENCE TESTS Nominal Flow Rate (ml per hour) High Limit (ml per hour) Low Limit (ml

Page 68 - External Resistance

medTester 5000C Operators Manual 1-4 Where to Find Help If you have questions not answered in this manual, please refer to the following Fluke Biomedi

Page 69

medTester 5000C Operators Manual 10-22 Customizing IV Pump Tests (For detailed information about customizing autosequences, please refer to Chapter 6

Page 70

Biomedical Test System IV Pump Module10 10-23 1. Name—This is the name you want to give to the autosequence, with a maximum of six characters. 2

Page 71

medTester 5000C Operators Manual 10-24 2. SECONDARY RATE HIGH LIMIT: Enter the high limit for the flow rate. 0000.0 turns off high limit checking. 3.

Page 72

Biomedical Test System IV Pump Module10 10-25 4. RATE LOW LIMIT: Enter the low limit for the flow rate. 0000.0 turns off low limit checking. 5.

Page 73 - A personal computer

medTester 5000C Operators Manual 10-26 PRINTF3 Reinitialize Factory Default Settings Any customized autosequences can be returned to the factory defau

Page 74

11-1 Chapter 11 medCheck Module In this chapter you will find information about checklists—what they are, how they are used, running them in the med

Page 75 - ENT (Enter) key

medTester 5000C Operators Manual 11-2 With medCheck, your medTester can interface and control external equipment such as the Impulse 4000 Defibrillato

Page 76 - • Peformance Comments

Biomedical Test System medCheck Module11 11-3 Understanding and Using Checklists Checklists let you standardize and quickly test equipment, accu

Page 77

medTester 5000C Operators Manual 11-4 ABC HOSPITAL Defibrillator Checklist Checkitem Description Control Code A5 A Chassis and housing…General condi

Page 78

Biomedical Test System medCheck Module11 11-5 As you can see, some checkitems appear in the medTester as prompts to the operator to do somethin

Page 79 - TURN EUT POWER OFF

Biomedical Test System General Information 1 1-5 Features The medTester 5000C is an automated biomedical equipment test system designed for you

Page 80 - OUTPUT PERFORMANCE WAVES?

medTester 5000C Operators Manual 11-6 You’ve Made a Checklist—Now What? While the process of making a checklist takes place in your CMMS, using it tak

Page 81

Biomedical Test System medCheck Module11 11-7 This message shows you that checklists have been loaded into the medTester 5000C. The characters X

Page 82 - • Op (operator) code

medTester 5000C Operators Manual 11-8 1. OP CODE—This is the operator code for the person doing the testing. This code can consist of a maximum of thr

Page 83 - Test Tag Printer

Biomedical Test System medCheck Module11 11-9 Note Turning off pretest prompts for checklists helps save time in executing checklists on severa

Page 84 - / TAG menu path

medTester 5000C Operators Manual 11-10 Input Requested (I) Category This checkitem tells the operator to check something or to do something and then i

Page 85 - CUSTOM/TAG/CONFIG menu

Biomedical Test System medCheck Module11 11-11 If an external command is followed by the number 0, the medTester will not expect or look for a r

Page 86 - PRINT if you

medTester 5000C Operators Manual 11-12 Posttest Prompts Posttest prompts that have been switched on appear upon completion of the checklist. After res

Page 87 - • USER TIME

Biomedical Test System medCheck Module11 11-13 Remote Control of Fluke Biomedical Testers You can have the medTester 5000C control other instrum

Page 88

medTester 5000C Operators Manual 11-14

Page 89 - Toolbox Monitors

12-1 Chapter 12 Remote Operation In this chapter you find information on both local and remote mode operation of the medTester 5000C. In addition,

Page 90

medTester 5000C Operators Manual 1-6 Prompts in the autosequence ask you for information that aids in test reporting. The pretest prompts, for example

Page 91

medTester 5000C Operators Manual 12-2 Additional COM2 Functions Two-way Communications—for commands to and responses from Fluke Biomedical testers and

Page 92

Biomedical Test System Remote Operation12 12-3 COM2 baud rates are selected from: COM2F2 Note If you have enabled the Wedge adapter, there will

Page 93 - Customize Your medTester

medTester 5000C Operators Manual 12-4 You can connect the optional RS-232 keyboard or laser bar code gun to either serial port. Input from this keyboa

Page 94

Biomedical Test System Remote Operation12 12-5 Record Output If you enter RECORDS, you see a display similar to this: Configuring Ports for Rec

Page 95 - 10LD X, the

medTester 5000C Operators Manual 12-6 Configuring Ports for Test Tag Output Configure any of these three ports for test tags as described above for re

Page 96 - Customizing Autosequence

Biomedical Test System Remote Operation12 12-7 Forcing A Port An alternative to the above procedure is forcing a port. For example, if you want

Page 97 - LEAKAGE CURRENT TESTS

medTester 5000C Operators Manual 12-8 UNIT of MEASUREMENT ABBREVIATIONS Abbreviation Unit of Measurement V RMS Volts RMS A RMS Amperes RMS µA RMS

Page 98

Biomedical Test System Remote Operation12 12-9 ERROR RESPONSES Error Statement Explanation the file has a name embedded into it, the wrong file

Page 99 - Test Record Header

medTester 5000C Operators Manual 12-10 Uploading Here is the file upload process: 1. The medTester receives an uploading command from the computer in

Page 100

Biomedical Test System Remote Operation12 12-11 Downloading The medTester 5000C receives file download commands from the remote computer in the

Page 101 - Turning Prompts On and Off

Biomedical Test System General Information 1 1-7 Checklists are complete sets of preventive maintenance procedures for equipment testing. They c

Page 102 - Renaming Prompts

medTester 5000C Operators Manual 12-12 Remote Commands List Commands sent to the medTester from a remote computer include: • General Administrative C

Page 103 - Test Tag Configuration

Biomedical Test System Remote Operation12 12-13 FILE TRANSFER COMMANDS (Note: Usage of these commands is described earlier.) Command Explanat

Page 104 - Summary of Stored Records

medTester 5000C Operators Manual 12-14 MEASUREMENT COMMANDS Note: These commands set up a measurement, according to global variables as needed, and

Page 105

Biomedical Test System Remote Operation12 12-15 MEASUREMENT COMMANDS (µA RMS or DC) LALD ECG leakage current, Left Arm to leads. (µA RMS or DC

Page 106 - F3 to accept

medTester 5000C Operators Manual 12-16 WAVE COMMANDS Note: The following commands turn on the specified wave. The wave continues to run until it is

Page 107 - • Chapter 15, SPO2 Module

Biomedical Test System Remote Operation12 12-17 Port Diagnostics You can test the input and output of COM1 and COM2, as well as the output of th

Page 108

medTester 5000C Operators Manual 12-18

Page 109 - Performance Waves

13-1 Chapter 13 The Wedge Adapter In this chapter you learn about the Wedge adapter and how to use it, how to attach it to the medTester 5000C, and

Page 110 - Waveform Groups

medTester 5000C Operators Manual 13-2 Serial Port Expansion By expanding the number of available ports on a medTester 5000C with the Wedge, you have t

Page 111 - WAVEFORM GROUPS

Biomedical Test System The Wedge Adapter13 13-3 When you execute a checklist, the medTester 5000C directs any M-type commands through the device

Page 112 - Esc twice to exit

Warranty and Product Support Fluke Biomedical warrants this instrument against defects in materials and workmanship for one full year from the date of

Page 113 - Autosequences

medTester 5000C Operators Manual 1-8 These modules are software enabled for use by the medTester’s firmware. The firmware is held in EPROM (erasable p

Page 114

medTester 5000C Operators Manual 13-4 Wedge Port Configurations Wedge Port Name (Default Wedge Port No.) Description Non-Invasive Blood Pressure (NIBP

Page 115 - Chapter 8

Biomedical Test System The Wedge Adapter13 13-5 Disassembly of the medTester 5000C’s Feet and Tilt Bail Disassemble the medTester 5000C feet an

Page 116 - Printing Records

medTester 5000C Operators Manual 13-6 FROM: Wedge Port (9-pin connection) TO: medTester 5000C (25-pin connection) MEDT COM1 COM1 MEDT COM2 COM2

Page 117 - RECORD #

Biomedical Test System The Wedge Adapter13 13-7 Operating the Wedge Operating the medTester 5000C with the Wedge consists of enabling the wedge,

Page 118

medTester 5000C Operators Manual 13-8 Configuring the Normal Port The Wedge normal port functions identically to the local input and output modes of t

Page 119 - Configuring Your Printer

Biomedical Test System The Wedge Adapter13 13-9 Configuring an Autosequence Port You can execute defibrillator, IV pump, ESU, and SPO2 autoseque

Page 120 - Checking Contents of Memory

medTester 5000C Operators Manual 13-10 Configure other autosequence ports as described here. Configuring the Checklist Port Whenever you execute a che

Page 121 - Searching for Records

Biomedical Test System The Wedge Adapter13 13-11 5. After selecting the normal port, press F5 to store your selection. Changing the Checklist Po

Page 122 - STRING NOT FOUND

medTester 5000C Operators Manual 13-12 Note Autosequences always communicate at a fixed baud rate; therefore, you don't need to set the rate for

Page 123 - Keyboard Shortcut Commands

Biomedical Test System The Wedge Adapter13 13-13 5. After setting the baud rate for the port, press F5 to store the setting. Setting Output Po

Page 124

Biomedical Test System General Information 1 1-9 medTester 5000C Modules medTester 5000C Module Name Functions Enabled medTester Prerequisites

Page 125 - Defibrillator Module

medTester 5000C Operators Manual 13-14

Page 126 - MODULE NOT INSTALLED

14-1 Chapter 14 ESU Module In this chapter you learn about using the medTester 5000C to control electro-surgery (ESU) autosequence tests with Fluke

Page 127

medTester 5000C Operators Manual 14-2 Using the ESU Module with the Model 402A and RF303RS ESU Analyzer The ESU module was originally designed to w

Page 128

Biomedical Test System ESU Module14 14-3 Note Refer to your ESU analyzer operators manual for additional information regarding operation and ran

Page 129 - NO, go to the prompt

medTester 5000C Operators Manual 14-4 ESU Autosequence Names You can begin any of the autosequences listed below from the medTester menu commands or d

Page 130 - ENT key

Biomedical Test System ESU Module14 14-5 Running ESU Autosequence Tests To begin running ESU autosequence tests, connect your medTester 5000C to

Page 131

medTester 5000C Operators Manual 14-6 To accept previously input prompt information, press YES, and go to the Test Sequence section. If you choose NO,

Page 132

Biomedical Test System ESU Module14 14-7 8. PHYSICAL INSPECTION—You have two physical inspection lines, each of which can consist of a maximum o

Page 133 - ENERGY = 100.X J

medTester 5000C Operators Manual 14-8 The setup screen prompts the user to set up the ESU for power setting, mode, and output. Push F5 READY when the

Page 134 - CONTINUE

Biomedical Test System ESU Module14 14-9 CQM (Contact Quality Monitor) Test The CQM test measures the resistance needed by the ESU device to act

Page 135 - Defibrillator Module

medTester 5000C Operators Manual 1-10 medTester 5000C Modules medTester 5000C Module Name Functions Enabled medTester Prerequisites for Use of Module

Page 136 - DEFIB menu, select:

medTester 5000C Operators Manual 14-10 ESU Autosequence Descriptions Autosequence Number: A60 Autosequence Name: ASPENX Step and Test Type Load ES

Page 137

Biomedical Test System ESU Module14 14-11 Autosequence Number: A61 Autosequence Name: VLF30 Step and Test Type Load ESU Mode Output Power

Page 138 - MENU1, press the

medTester 5000C Operators Manual 14-12 Autosequence Number: A62 Autosequence Name: VLF40 Step and Test Type Load ESU Mode Output Power Setting E

Page 139 - IV Pump Module

Biomedical Test System ESU Module14 14-13 Autosequence Number: A63 Autosequence Name: VLF4BF Step and Test Type Load ESU Mode Output Power

Page 140 - IDA 4 Plus

medTester 5000C Operators Manual 14-14 Autosequence Number: A64 Autosequence Name: VLSSE4 Step and Test Type Load ESU Mode Output) Power Setting

Page 141

Biomedical Test System ESU Module14 14-15 Customizing ESU Autosequences (For detailed information about customizing autosequences, please refer

Page 142

medTester 5000C Operators Manual 14-16 Begin by accessing the custom, autosequence, ESU menu as described above. This time, select: MAKEF1 Next, selec

Page 143 - IV Pump Autosequence Names

Biomedical Test System ESU Module14 14-17 • Expected Power: Enter the power expected to be generated by the ESU, from 0 to 500 W. • Power hig

Page 144

medTester 5000C Operators Manual 14-18 STOREF5 Print Your Autosequence Printing any autosequence is accomplished through this menu command: PRINTF3 Us

Page 145 - IV Pump Module

15-1 Chapter 15 SPO2 Module In this chapter you learn about using the medTester 5000C to control SPO2 autosequence tests with Fluke Biomedical SPO2

Page 146

Biomedical Test System General Information 1 1-11 Features • Fully automated electrical safety testing • 12–lead ECG/arrhythmia simulation •

Page 147

medTester 5000C Operators Manual 15-2 PLUS 7), the CardioSat 100 or Index 2XL. All autosequences stored by initialization, customization, or remote do

Page 148 - IV PUMP TESTS

Biomedical Test System SPO2 Module15 15-3 If SPO2 autosequences are installed, you see this message displayed: The above display format an

Page 149 - Dual Rate Test

medTester 5000C Operators Manual 15-4 SPO2 Autosequence Names You can begin any of the autosequences listed below from the medTester menu commands or

Page 150 - PCA Test

Biomedical Test System SPO2 Module15 15-5 SP02 AUTOSEQUENCES FOR INDEX 2XL Autosequence Name Keyboard Shortcut Command Menu Commands from SP

Page 151 - 10-13

medTester 5000C Operators Manual 15-6 SP02 AUTOSEQUENCES FOR CARDIOSAT 100 Autosequence Name Keyboard Shortcut Command Menu Commands from SPO2F5AUT

Page 152

Biomedical Test System SPO2 Module15 15-7 Note The Index 2XL must have been manufactured after 5/21/04 or modified by Fluke Biomedical (ECO 416

Page 153

medTester 5000C Operators Manual 15-8 Pretest Device Prompts There are eight prompts for information that you can attach to test records. After select

Page 154

Biomedical Test System SPO2 Module15 15-9 Note See Chapter 5, Autosequences, for a complete explanation of working with prompts. You can turn i

Page 155 - 10-17

medTester 5000C Operators Manual 15-10 Each autosequence can contain up to ten SPO2 simulation steps, except the INDEX 2XL which can contain up to 15

Page 156 - @ sign (with a

Biomedical Test System SPO2 Module15 15-11 You should now observe the readings on the pulse oximeter for SPO2 and heart rate. You can enter tes

Page 157

medTester 5000C Operators Manual 1-12 medTester 5000C Instrument Specifications Line Voltage and Measurements • Hot to Neutral. • Neutral to Ground.

Page 158 - IPT-1 AUTOSEQUENCE TESTS

medTester 5000C Operators Manual 15-12 Probe Test For the Oxitest PLUS, there is no probe test built-in. However, it is still possible to include a pr

Page 159 - IPT-MC AUTOSEQUENCE TESTS

Biomedical Test System SPO2 Module15 15-13 For the Oxitest PLUS: SPO2 Autosequence Number: A70 Autosequence Name: ADULT1 Settling Time: 25

Page 160 - Customizing IV Pump Tests

medTester 5000C Operators Manual 15-14 For the Oxitest PLUS: SPO2 Autosequence Number: A72 Autosequence Name: ADULT3 Settling Time: 25 Beats Heart

Page 161

Biomedical Test System SPO2 Module15 15-15 For the Oxitest PLUS: SPO2 Autosequence Number: A74 Autosequence Name: NEO2 Settling Time: 25 Bea

Page 162

medTester 5000C Operators Manual 15-16 For CardioSat 100: SPO2 Autosequence Number: A70 Autosequence Name: NELLCR Manufacturer: Nellcor Settling Ti

Page 163

Biomedical Test System SPO2 Module15 15-17 For CardioSat 100: SPO2 Autosequence Number: A72 Autosequence Name: HPMRLN Manufacturer: Hewlett

Page 164 - MENU 2

medTester 5000C Operators Manual 15-18 For CardioSat 100: SPO2 Autosequence Number: A74 Autosequence Name: GENERC Manufacturer: Generic–Not Specifi

Page 165 - Chapter 11

Biomedical Test System SPO2 Module15 15-19 For the Index 2XL: SPO2 Autosequence Number: A70 Autosequence Name: NELLCR Manufacturer: Nellcor

Page 166 - What’s a Checklist?

medTester 5000C Operators Manual 15-20 For the Index 2XL: SPO2 Autosequence Number: A72 Autosequence Name: PMS Manufacturer: PMS M1190 Settling Ti

Page 167 - medCheck Module

Biomedical Test System SPO2 Module15 15-21 For the Index 2XL: SPO2 Autosequence Number: A74 Autosequence Name: GENERC Manufacturer: Generic

Page 168 - Defibrillator Checklist

Biomedical Test System General Information 1 1-13 AAMI Load • Simulated patient load recommended by the Association for the Advancement of Medi

Page 169

medTester 5000C Operators Manual 15-22 Posttest Prompts When all tests are completed, the medTester 5000C prompts you for: 1. COMMENTS 2. NEXT TEST

Page 170

Biomedical Test System SPO2 Module15 15-23 Note You can find detailed information about customizing autosequences in Chapter 6, Customize Your

Page 171

medTester 5000C Operators Manual 15-24 Not Specified BCI Criticare Datascope Datex Hewlett Packard Nellcor Nihon Kohden Novametrix Ohmeda Respironics

Page 172

Biomedical Test System SPO2 Module15 15-25 Note If you choose "Not Specified", you can keep it generic. However, you must manually se

Page 173 - Pass/Fail (P) Category

medTester 5000C Operators Manual 15-26 MOREF4 to select more steps, or ENDF5 to end your selections. 6. STARTING ALARM VALUES: Enter starting value

Page 174

Biomedical Test System SPO2 Module15 15-27 Print Your Autosequence From the SPO2 custom autosequences menu described above, printing any autose

Page 175 - 10 SECOND DELAY

medTester 5000C Operators Manual 15-28

Page 176 - Checklist Records

16-1 Chapter 16 Transcutaneous Pacemaker (Pacer) Module In this chapter you learn about using the medTester 5000C to control transcutaneous pacemake

Page 177 - 11-13

medTester 5000C Operators Manual 16-2 For related information, see these chapters in this manual: • Chapter 5, Autosequences. • Chapter 6, Customize

Page 178

Biomedical Test System Transcutaneous Pacemaker (Pacer) Module16 16-3 Pacer Autosequence Names You can begin any of the autosequences listed bel

Page 179 - Remote Operation

medTester 5000C Operators Manual 1-14 Toolbox Toolbox allows for the use of external measurement adapters for: • Tachometer—with a range of 100 to 20

Page 180 - Setting Baud Rates

medTester 5000C Operators Manual 16-4 Running Pacer Autosequence Tests To begin running pacer autosequence tests, connect your medTester 5000C to your

Page 181 - Local Input Mode

Biomedical Test System Transcutaneous Pacemaker (Pacer) Module16 16-5 Pretest Device Prompts There are eight prompts for information that can ap

Page 182

medTester 5000C Operators Manual 16-6 Note See Chapter 5, Autosequences, for a complete explanation of working with prompts. You can turn individual p

Page 183 - Remote Operation

Biomedical Test System Transcutaneous Pacemaker (Pacer) Module16 16-7 Pulse Tests For each of the pulse test steps, 1 through 10, the medTester

Page 184 - Remote Mode

medTester 5000C Operators Manual 16-8 Demand Test The demand test checks that in the demand mode, the pacer will pulse interactively with ECG waveform

Page 185 - Forcing A Port

Biomedical Test System Transcutaneous Pacemaker (Pacer) Module16 16-9 Async Test The async test checks that in the asynchronous, or non-demand,

Page 186 - ERR=xx. Here is a list of

medTester 5000C Operators Manual 16-10 Pacer Autosequence Number: A20 Autosequence Name: LP12P Load: Internal 50 Ohms Step Number Current (in mill

Page 187

Biomedical Test System Transcutaneous Pacemaker (Pacer) Module16 16-11 Pacer Autosequence Number: A21 Autosequence Name: LP9PM Load: Adapter

Page 188 - ERR=24,LINE COUNT WRONG

medTester 5000C Operators Manual 16-12 Pacer Autosequence Number: A22 Autosequence Name: HPCMPM Load: Adapter Module TQA-12, External 400 Ohms Step

Page 189 - ABORT message

Biomedical Test System Transcutaneous Pacemaker (Pacer) Module16 16-13 Pacer Autosequence Number: A23 Autosequence Name: MDE3PM Load: Adapter

Page 190 - Remote Commands List

Biomedical Test System General Information 1 1-15 Ground Fault Circuit Interrupter The test receptacle is protected by a ground fault circuit in

Page 191

medTester 5000C Operators Manual 16-14 Pacer Autosequence Number: A24 Autosequence Name: ZOLLPM Load: Adapter Module TQA-2, External 1000 Ohms Step

Page 192

Biomedical Test System Transcutaneous Pacemaker (Pacer) Module16 16-15 Customizing Pacer Autosequences (For detailed information about customizi

Page 193

medTester 5000C Operators Manual 16-16 Begin by accessing the custom, autosequence, pacer menu as described earlier. This time, select: MAKEF1 Next se

Page 194 - Command Explanation

Biomedical Test System Transcutaneous Pacemaker (Pacer) Module16 16-17 Reinitialize Factory Default Settings Any customized autosequences can be

Page 195 - Port Diagnostics

medTester 5000C Operators Manual 16-18

Page 196

17-1 Chapter 17 Non-Invasive Blood Pressure (NIBP) Module In this chapter you learn about using the medTester 5000C to control non-invasive blood pr

Page 197 - The Wedge Adapter

medTester 5000C Operators Manual 17-2 For related information, see these chapters in this manual: • Chapter 5, Autosequences. • Chapter 6, Customize

Page 198 - Port Configuration

Biomedical Test System Non-Invasive Blood Pressure (NIBP) Module17 17-3 NIBP Autosequence Names You can begin any of the autosequences listed be

Page 199

medTester 5000C Operators Manual 17-4 Running NIBP Autosequence Tests To begin running NIBP autosequence tests, connect your medTester 5000C to your C

Page 200 - Installing the Wedge

Biomedical Test System Non-Invasive Blood Pressure (NIBP) Module17 17-5 Pretest Device Prompts There are eight prompts for information that can

Page 201

medTester 5000C Operators Manual 1-16 General Specifications Parameter Specification Power Requirement: 115 V ±10%, 60 Hz only Temperature Ranges:Oper

Page 202 - MEDT COM2 COM2

medTester 5000C Operators Manual 17-6 Note See Chapter 5, Autosequences, for a complete explanation of working with prompts. You can turn individual p

Page 203 - Operating the Wedge

Biomedical Test System Non-Invasive Blood Pressure (NIBP) Module17 17-7 The medTester records the following test parameters from the CuffLink fo

Page 204 - F5 to store your selection

medTester 5000C Operators Manual 17-8 PRESSURE TEST The pressure test allows you to document up to three static pressure levels generated by the CuffL

Page 205

Biomedical Test System Non-Invasive Blood Pressure (NIBP) Module17 17-9 LEAK TEST The leak test measures the rate of leakage in the NIBP measure

Page 206

medTester 5000C Operators Manual 17-10 NIBP Autosequence Number: A80 Autosequence Name: ADULT1 Step Number Blood Pressure Heart Rate in BPM (beats

Page 207

Biomedical Test System Non-Invasive Blood Pressure (NIBP) Module17 17-11 NIBP Autosequence Number: A83 Autosequence Name: NEONT1 Step Number

Page 208

medTester 5000C Operators Manual 17-12 Customizing NIBP Autosequences (For detailed information about customizing autosequences, please refer to Chapt

Page 209 - Setting Output Ports

Biomedical Test System Non-Invasive Blood Pressure (NIBP) Module17 17-13 Begin by accessing the custom, autosequence, NIBP menu as described ear

Page 210

medTester 5000C Operators Manual 17-14 After customizing your autosequence, the medTester prompts you to store the autosequence: STOREF5 Print Your Au

Page 211 - ESU Module

Biomedical Test System Index Index-1 Index A AAMI load, 1-13 Accessories, 1-17 Audio Transducer, 2-4 Autosequences defined, 5-1 monitoring envi

Page 212 - • Chapter 5, Autosequences

Biomedical Test System General Information 1 1-17 Accessories Standard Part # • 20/15–amp Adapter 2195732 • Two Kelvin Cable 2392617 • Two

Page 213

Index-2 checklists, using, 11-3 Memory checking contents of, 8-6 Menus navigating, 3-4 Modules, list of, 1-8 modules, optional overview, 1-2 Modules,

Page 214

Biomedical Test System Index Index-3 installing, 13-4 operating, 13-7 setting baud rates, 13-11 setting output ports, 13-13

Page 215 - ESU Module

medTester 5000C Operators Manual Index-4

Page 216 - YES, and go to the Test

Notices All Rights Reserved  Copyright 2006, Fluke Biomedical. No part of this publication may be reproduced, transmitted, transcribed, stored in a r

Page 217

medTester 5000C Operators Manual 1-18 Optional (continued) Part # • Printer Cable, medTester to standard PC printer Serial Cable; Male DB25 to Cent

Page 218 - RF Leakage

Biomedical Test System General Information 1 1-19 medTester 5000C Module Upgrades Part #Factory Installed • Module 2, Service, RS-232/Printe

Page 219

medTester 5000C Operators Manual 1-20

Page 220

2-1 Chapter 2 Installation In this chapter you will find information about the default (factory) settings, power-up sequence, setting the time and d

Page 221 - 14-11

medTester 5000C Operators Manual 2-2 FACTORY DEFAULT SETTINGS Condition Name Default Setting(s) Output • RECORD and TAG output set to: • COM1–OFF •

Page 222

Biomedical Test System Installation 2 2-3 medTester 5000C Power-Up Condition Messages Message Condition Indicated LOW BATTERY, SERVICE REQUIRED

Page 223 - 14-13

medTester 5000C Operators Manual 2-4 4. After entering date and time, insert a thin, non-metallic device in the “clock set enable switch” slot (labele

Page 224

Biomedical Test System Installation 2 2-5 Enabling Modules If it should become necessary to reinstall modules 2 through 14, perform the steps de

Page 225 - menu, select:

medTester 5000C Operators Manual 2-6 Note If running Windows 95 or 98, with the mouse you can click and hold the Start button, then select Programs an

Page 226

Biomedical Test System Installation 2 2-7 10. To install more than one module, repeat steps seven (7) through nine (9) for each module which has

Page 227

Restocking Charges Products returned within 30 days of original purchase are subject to a minimum restocking fee of 15 %. Products returned in excess

Page 228

medTester 5000C Operators Manual 2-8

Page 229 - SPO2 Module

3-1 Chapter 3 Instrument Familiarity This chapter familiarizes you with the medTester 5000C. Know Your medTester 5000C The medTester 5000C has two m

Page 230

medTester 5000C Operators Manual 3-2 MED TESTER 500011413 15 16 17 18 1923851246791011 Figure 3-1. MedTester 5000C Top and Rear Panel Locator

Page 231 - SPO2 Module

Biomedical Test System Instrument Familiarity 3 3-3 6. ARROW KEYS—Control cursor movement when entering data into the display from the keyboar

Page 232

medTester 5000C Operators Manual 3-4 16. COM1 PORT—Receives input locally from the optional keyboard, bar code scanning gun, or personal computer or t

Page 233

Biomedical Test System Instrument Familiarity 3 3-5 The display now shows menu two. Remember that menu two is an extension of menu one. Menu-tw

Page 234

medTester 5000C Operators Manual 3-6

Page 235

4-1 Chapter 4 Manual Tests Learn to use the medTester 5000C to manually perform electrical safety tests for line voltage, leakage current, equipment

Page 236

medTester 5000C Operators Manual 4-2 Hot to Neutral Enter menu commands: MANUALF2VOLTSF1HOT-NEUTF1 Neutral to Ground Enter menu commands: MANUALF2VOLT

Page 237

Biomedical Test System Manual Tests 4 4-3 External Case Leakage See Figure 4-2, Case External Leakage. 1. Plug the EUT (equipment under test) i

Page 238 - LEVEL 2

traceable to the National Institute of Standards and Technology (NIST). Devices for which there are no NIST calibration standards are measured against

Page 239

medTester 5000C Operators Manual 4-4 • POLARITY—normal or reversed, • NEUTRAL—open or closed. The respective condition LEDs light, and the display

Page 240 - DOWN/UP keys, then

Biomedical Test System Manual Tests 4 4-5 • Lead isolation, all leads. • Lead isolation, single lead Note For all lead-to-ground and lead-to-l

Page 241 - For the Oxitest PLUS:

medTester 5000C Operators Manual 4-6 RL—right leg, RA—right arm, LA—left arm, LL—left leg, and V—V1 through V6 leads tied together. You can measure V

Page 242

Biomedical Test System Manual Tests 4 4-7 Isolation Lead Leakage See Figure 4-7, ECG Lead Isolation Test: All Leads and Individual Leads. You c

Page 243 - For the Oxitest PLUS:

medTester 5000C Operators Manual 4-8 4. Select the keyboard function key that corresponds to the lead you wish to test. 5. To begin the measurement,

Page 244 - For CardioSat 100:

Biomedical Test System Manual Tests 4 4-9 Power Cord Resistance See Figure 4-9, Power Cord Resistance Test. 1. Plug the equipment under test in

Page 245

medTester 5000C Operators Manual 4-10 External Resistance See Figure 4-10, External Resistance. 1. Connect the dual banana-plug end of one Kelvin cab

Page 246

Biomedical Test System Manual Tests 4 4-11 Toolbox Toolbox consists of measurements that require the use of three adapters. These adapters, avai

Page 247 - For the Index 2XL:

medTester 5000C Operators Manual 4-12 Temperature 1. Plug the temperature adapter test leads into the EXT INPUT jacks on medTester’s front panel. 2.

Page 248

Biomedical Test System Manual Tests 4 4-13 Figure 4-1. Three Types of Line Voltage Measurements

Page 250

medTester 5000C Operators Manual 4-14 Figure 4-2. Case External Leakage

Page 251 - CARDIOSAT 100

Biomedical Test System Manual Tests 4 4-15 Figure 4-3. Case Internal Leakage

Page 252 - ANUFACTURER BASE TLC

medTester 5000C Operators Manual 4-16 Figure 4-4. External Leakage Current

Page 253

Biomedical Test System Manual Tests 4 4-17 Figure 4-5. ECG Lead Leakage to Ground: All Leads to Ground and Individual Lead to Ground

Page 254

medTester 5000C Operators Manual 4-18 Figure 4-6. ECG Interlead Leakage

Page 255

Biomedical Test System Manual Tests 4 4-19 Figure 4-7. ECG Lead Isolation Test: All Leads and Individual Leads

Page 256

medTester 5000C Operators Manual 4-20 Figure 4-8. Equipment Load Current

Page 257 - (Pacer) Module

Biomedical Test System Manual Tests 4 4-21 Figure 4-9. Power Cord Resistance Test

Page 258

medTester 5000C Operators Manual 4-22 Figure 4-10. External Resistance

Page 259

Biomedical Test System Manual Tests 4 4-23 Figure 4-11. Isolated Power System and Ground Fault Test

Page 260

Biomedical Test System Contents Table of Contents 1: GENERAL INFORMATION SAFETY CONSIDERATIONS ...

Page 261

medTester 5000C Operators Manual 4-24

Page 262

5-1 Chapter 5 Autosequences This chapter explains how to use the medTester 5000C’s electrical safety autosequences, two standard monitoring sequence

Page 263 - Pulse Tests

medTester 5000C Operators Manual 5-2 • Defibrillators—Using Fluke Biomedical defibrillator analyzers Impulse 3000 and Impulse 4000 (see Chapter 9, De

Page 264 - Demand Test

Biomedical Test System Autosequences 5 5-3 Note In the standard shipping configuration, tests are performed both with power turned on and with p

Page 265

medTester 5000C Operators Manual 5-4 Standard Safety Autosequence Names You can access autosequences from either medTester menu commands or from the k

Page 266

Biomedical Test System Autosequences 5 5-5 SAFETY AUTOSEQUENCES TEMPMON Temperature Monitor TOOLBOXF2TEMPMONF4 A18 HUMMON Humidity Monitor TOOL

Page 267 - 16-11

medTester 5000C Operators Manual 5-6 Note All these pre-test prompts can be turned off. However, if using the medTester with an Equipment Management S

Page 268

Biomedical Test System Autosequences 5 5-7 Autosequence Steps If you have autosequence prompts turned on, you enter the appropriate information.

Page 269 - 16-13

medTester 5000C Operators Manual 5-8 System Line Voltage Measurements included are: • Hot to neutral, • Neutral to ground, and • Hot to ground. Pow

Page 270

Biomedical Test System Autosequences 5 5-9 Leakage Current Measurements After the power cord resistance test, the autosequence moves to leakage

Page 271 - PACER menu, select:

Date and Time Setup ... 2-3 Audio Transducer...

Page 272

medTester 5000C Operators Manual 5-10 ECG Performance Waveforms If you have the Waveform/Extended Testing module installed on your medTester 5000C, an

Page 273

Biomedical Test System Autosequences 5 5-11 Note See Chapter 7, Performance Waves, for detailed instructions about performance waveform output.

Page 274

medTester 5000C Operators Manual 5-12 This prompt appears only if you enable the prompt in the CUSTOM menu. See Chapter 6, Customize Your medTester 50

Page 275 - (NIBP) Module

Biomedical Test System Autosequences 5 5-13 Test Tag Printer After performing an autosequence or checklist (test procedure created in your CMMS)

Page 276

medTester 5000C Operators Manual 5-14 Note To enable or disable any of the five test tag fields (lines), see Test Tag Configuration in Chapter 6, Cus

Page 277

Biomedical Test System Autosequences 5 5-15 1. From the menu, press F1 (CONFIG) as before to configure your test tag (test tag lines on/off), or

Page 278

medTester 5000C Operators Manual 5-16 Menu Selection Procedure Both monitoring autosequences are found under AUTOF1. Press the right arrow key on the

Page 279

Biomedical Test System Autosequences 5 5-17 1. OP CODE—The operator is the person doing the testing. This code can consist of a maximum of three

Page 280 - NIBP TESTS

medTester 5000C Operators Manual 5-18 After you select ENVMONF3 from the menu, medTester prompts— Select MEMORY if you want to store the test record.

Page 281

Biomedical Test System Autosequences 5 5-19 End-of-Test Prompts At the end of the autosequence, medTester prompts you for the following data: (p

Page 282 - PRESSURE TEST

Biomedical Test System Contents What You See in an Autosequence...6-2 How to Customize...

Page 283

medTester 5000C Operators Manual 5-20 4. For resetting the maximum reading to become the minimum, press the RESET (F1) key. The medTester will prompt

Page 284

Biomedical Test System Autosequences 5 5-21 Humidity Monitor You can monitor the level of relative humidity over periods of time, print the resu

Page 285

medTester 5000C Operators Manual 5-22 4. CYCLE TEST TIME IN MINUTES—Enter the time between measurements (1-9 minutes). 5. TRIGGER LEVEL—Enter the tri

Page 286 - NIBP menu, select:

6-1 Chapter 6 Customize Your medTester In this chapter you learn how to customize safety autosequences and how to customize the prompt information t

Page 287

medTester 5000C Operators Manual 6-2 5. After pressing this function key, you can begin scrolling through the steps in the sequence by using your up a

Page 288

Biomedical Test System Customize Your medTester 6 6-3 You can press Esc to move out of view and back to the previous menu choices, or you can pr

Page 289 - Index-1

medTester 5000C Operators Manual 6-4 Customizing Menu Steps As you step through the customizing menu, the medTester asks you to enter information. Not

Page 290 - Index-2

Biomedical Test System Customize Your medTester 6 6-5 LEAKAGE CURRENT TESTS Test Item Legend rev—reverse pol—polarity clsd—closed gnd—ground neu

Page 291 - Index-3

medTester 5000C Operators Manual 6-6 LEAKAGE CURRENT TESTS Test Item Legend rev—reverse pol—polarity clsd—closed gnd—ground neu—neutral iso—isolated n

Page 292 - Index-4

Biomedical Test System Customize Your medTester 6 6-7 Customizing a Blank Safety Autosequence If your medTester 5000C contains the Waveform/Exte