medTester 5000C Biomedical Tester Operators Manual PN 2243153 February 2006 2006 Fluke Corporation, All rights reserved. Printed in USA All product
Test Sequence... 9-7 Posttest Prompts...
medTester 5000C Operators Manual 6-8 Customizing Your Autosequence Prompts In an autosequence, the medTester 5000C prompts you for information that is
Biomedical Test System Customize Your medTester 6 6-9 medTester Autosequence Prompts Prompt # Prompt Name/ Renameable? (Yes or No) Description
medTester 5000C Operators Manual 6-10 Select OFF, or ON. Then press NEXT to move to prompt two. If this is the only prompt or the last prompt that you
Biomedical Test System Customize Your medTester 6 6-11 Pausing Between tests in a safety autosequence, you can insert a pause for three, six, o
medTester 5000C Operators Manual 6-12 Any or all of these fields can be turned on (to appear on the test tag) or turned off on the printed test tag. W
Biomedical Test System Customize Your medTester 6 6-13 RECSUMF1CUSTOMF3 The medTester displays the number of the first field to appear on each
medTester 5000C Operators Manual 6-14 Follow directions, such as cycling power after reprogramming the scanner. The scanner remains programmed after c
Biomedical Test System Customize Your medTester 6 6-15 Resetting Autosequences to Default Settings You can always return medTester 5000C autoseq
medTester 5000C Operators Manual 6-16
7-1 Chapter 7 Performance Waves In this chapter, you learn how to output performance waveforms to test patient monitors using the medTester 5000C. O
Biomedical Test System Contents Local Input Mode ...12-3 Local Output for Re
medTester 5000C Operators Manual 7-2 Waveform Groups You can generate any of the waveforms listed in the five groups in the table below for the testin
Biomedical Test System Performance Waves 7 7-3 WAVEFORM GROUPS medTester 5000C Display Wave Menu Name Description Remote Command PAIR A Pair o
medTester 5000C Operators Manual 7-4 Running Waveforms from The WAVES Menu To turn on any of the waveforms, follow these instructions: 1. Attach the
Biomedical Test System Performance Waves 7 7-5 Conducting Waveform Tests in Safety Autosequences You can execute performance wave tests at the e
medTester 5000C Operators Manual 7-6
8-1 Chapter 8 Memory This chapter acquaints you with the medTester 5000C memory module options you may have installed on your medTester. Introductio
medTester 5000C Operators Manual 8-2 Note All medTester memory functions begin on MENU2 as described above. Viewing A Single Record You can view reco
Biomedical Test System Memory 8 8-3 Print A Range of Records 1. Enter the menu command: RANGEF2 2. The display prompts you to PRINT FROM RECO
medTester 5000C Operators Manual 8-4 Next access the type of records you want to print. From menu one, you can access these types: • SAFETYF1 Prints
Biomedical Test System Memory 8 8-5 Printing A Summary of Records You can print a summary of all records stored from the menu command: SUMMARYF5
For the Oxitest PLUS: ... 15-13 For CardioSat 100: ...
medTester 5000C Operators Manual 8-6 Deleting Records If you want, you can delete a single record with this command: DELETEF3 The medTester asks you t
Biomedical Test System Memory 8 8-7 1. Free Memory—In bytes. 2. Total Memory and Checklists—Amount of total memory and the amount used by check
medTester 5000C Operators Manual 8-8 That indicates the number of the first record which contains the string. You can view or print the data in the fo
Biomedical Test System Memory 8 8-9 Keyboard Shortcut Commands In addition to accessing memory functions through menu commands, you can also ent
medTester 5000C Operators Manual 8-10
9-1 Chapter 9 Defibrillator Module In this chapter you learn about using the medTester 5000C to control defibrillator autosequence tests with Fluke
medTester 5000C Operators Manual 9-2 Note Consult your Impulse 3000 or Impulse 4000 operators manual for proper operation of your Impulse model. To di
Biomedical Test System Defibrillator Module 9 9-3 Defibrillator Autosequence Names You can begin any of the autosequences listed below from the
medTester 5000C Operators Manual 9-4 DEFIBRILLATOR AUTOSEQUENCES Autosequence Name Keyboard Shortcut Command Menu Commands from DEFIBF2AUTOF1 M/D2J
Biomedical Test System Defibrillator Module 9 9-5 Running Defibrillator Autosequence Tests To begin running defibrillator autosequence tests, co
1-1 Chapter 1 General Information In this chapter you will learn how to use this manual, where to get help, and about medTester 5000C Features and
medTester 5000C Operators Manual 9-6 Pretest Device Prompts There are eight prompts for information that can appear in the autosequence test record. A
Biomedical Test System Defibrillator Module 9 9-7 Note See Chapter 5, Autosequences, for a complete explanation of working with prompts. You can
medTester 5000C Operators Manual 9-8 Sternum Paddle. The sternum paddle test is similar to the APEX paddle test. Two (2) leakage measurements are take
Biomedical Test System Defibrillator Module 9 9-9 Out-of-Limits Reading When a reading is out of limits, a # sign appears next to the reading wi
medTester 5000C Operators Manual 9-10 MAX Energy Test. This test measures the maximum length of time the EUT requires to charge to its maximum energy
Biomedical Test System Defibrillator Module 9 9-11 The table below supplies specific information about the tests for each autosequence. Autosequ
medTester 5000C Operators Manual 9-12 Posttest Prompts When all tests are completed, the medTester 5000C can prompt you for: 1. COMMENTS 2. NEXT TES
Biomedical Test System Defibrillator Module 9 9-13 Begin by accessing the custom, autosequence, defib menu as described earlier. This time, sele
medTester 5000C Operators Manual 9-14 Reinitialize Factory Default Settings Any customized autosequences can be returned to the factory default settin
10-1 Chapter 10 IV Pump Module In this chapter you learn about using the medTester 5000C to control IV pump autosequence tests with a Fluke Biomedic
medTester 5000C Operators Manual 1-2 Introduction This manual is written for the biomedical technician or clinical engineer responsible for testing ho
medTester 5000C Operators Manual 10-2 Use the CUSTOM / AUTOSEQ / IVPUMP / TYPE menu path to configure the medTester for a specific IV pump analyzer.
Biomedical Test System IV Pump Module10 10-3 Infutest 2000 Series D The Infutest can run three types tests: single rate tests, dual rate tests,
medTester 5000C Operators Manual 10-4 Note In order to take advantage of IV pump autosequences, your medTester 5000C requires the IV Pump Module, RS-
Biomedical Test System IV Pump Module10 10-5 The above display format and testing protocol is formatted for the Infutest 2000 Series D IV Pump A
medTester 5000C Operators Manual 10-6 IV Pump Autosequences for IPT-1 Autosequence Name (Type of Test) Keyboard Shortcut Command Menu Commands from
Biomedical Test System IV Pump Module10 10-7 IV Pump Autosequences for IPT-MC Autosequence Name (Type of Test) Keyboard Shortcut Command Menu C
medTester 5000C Operators Manual 10-8 Running IV Pump Autosequences To begin running IV pump autosequence tests, connect your medTester 5000C to your
Biomedical Test System IV Pump Module10 10-9 1. OP CODE—This is the operator code for the person doing the testing. This code can consist of a m
medTester 5000C Operators Manual 10-10 IV PUMP TESTS After you respond to prompts, the autosequence tests begin. For each step in the autosequence, th
Biomedical Test System IV Pump Module10 10-11 Dual Rate Test • The dual rate test measures the delivery of two sequential fluid volumes pumped
Biomedical Test System General Information 1 1-3 The medTester 5000C—reliable safety and performance testing. Capable of managing your entire eq
medTester 5000C Operators Manual 10-12 • The medTester displays test results on 3 screens. Press F5 NEXT to go to the next screen. Finally, you can r
Biomedical Test System IV Pump Module10 10-13 IPT-1 • With the IPT-1, you can perform 2 types of flow tests: steady flow and high accuracy. Use
medTester 5000C Operators Manual 10-14 • For each test step, the medTester prompts you with the test type and conditions. Press F5 NEXT to continue.
Biomedical Test System IV Pump Module10 10-15 • Wait for the IDA 4 Plus to display PRIME on channel 1. Using a priming syringe, prime IV line i
medTester 5000C Operators Manual 10-16 • In a dual rate test, the medTester starts testing the secondary rate, indicated by “SEC” in the upper right
Biomedical Test System IV Pump Module10 10-17 • The IDA 4 Plus will measure delivered bolus volumes ≥ 0.5 ml with no basal (or background) flui
medTester 5000C Operators Manual 10-18 Pressure Test (Infutest, IPT-1, IPT-MC and IDA 4 Plus) • After completing the flow test, the autosequence cont
Biomedical Test System IV Pump Module10 10-19 INFUTEST 2000 Series D AUTOSEQUENCE TESTS Nominal Flow Rate (ml per hour) Test Type High Limit (
medTester 5000C Operators Manual 10-20 IPT-1 AUTOSEQUENCE TESTS Nominal Flow Rate (ml per hour) Test Type High Limit (ml per hour) Low Limit (ml per
Biomedical Test System IV Pump Module10 10-21 IPT-MC AUTOSEQUENCE TESTS Nominal Flow Rate (ml per hour) High Limit (ml per hour) Low Limit (ml
medTester 5000C Operators Manual 1-4 Where to Find Help If you have questions not answered in this manual, please refer to the following Fluke Biomedi
medTester 5000C Operators Manual 10-22 Customizing IV Pump Tests (For detailed information about customizing autosequences, please refer to Chapter 6
Biomedical Test System IV Pump Module10 10-23 1. Name—This is the name you want to give to the autosequence, with a maximum of six characters. 2
medTester 5000C Operators Manual 10-24 2. SECONDARY RATE HIGH LIMIT: Enter the high limit for the flow rate. 0000.0 turns off high limit checking. 3.
Biomedical Test System IV Pump Module10 10-25 4. RATE LOW LIMIT: Enter the low limit for the flow rate. 0000.0 turns off low limit checking. 5.
medTester 5000C Operators Manual 10-26 PRINTF3 Reinitialize Factory Default Settings Any customized autosequences can be returned to the factory defau
11-1 Chapter 11 medCheck Module In this chapter you will find information about checklists—what they are, how they are used, running them in the med
medTester 5000C Operators Manual 11-2 With medCheck, your medTester can interface and control external equipment such as the Impulse 4000 Defibrillato
Biomedical Test System medCheck Module11 11-3 Understanding and Using Checklists Checklists let you standardize and quickly test equipment, accu
medTester 5000C Operators Manual 11-4 ABC HOSPITAL Defibrillator Checklist Checkitem Description Control Code A5 A Chassis and housing…General condi
Biomedical Test System medCheck Module11 11-5 As you can see, some checkitems appear in the medTester as prompts to the operator to do somethin
Biomedical Test System General Information 1 1-5 Features The medTester 5000C is an automated biomedical equipment test system designed for you
medTester 5000C Operators Manual 11-6 You’ve Made a Checklist—Now What? While the process of making a checklist takes place in your CMMS, using it tak
Biomedical Test System medCheck Module11 11-7 This message shows you that checklists have been loaded into the medTester 5000C. The characters X
medTester 5000C Operators Manual 11-8 1. OP CODE—This is the operator code for the person doing the testing. This code can consist of a maximum of thr
Biomedical Test System medCheck Module11 11-9 Note Turning off pretest prompts for checklists helps save time in executing checklists on severa
medTester 5000C Operators Manual 11-10 Input Requested (I) Category This checkitem tells the operator to check something or to do something and then i
Biomedical Test System medCheck Module11 11-11 If an external command is followed by the number 0, the medTester will not expect or look for a r
medTester 5000C Operators Manual 11-12 Posttest Prompts Posttest prompts that have been switched on appear upon completion of the checklist. After res
Biomedical Test System medCheck Module11 11-13 Remote Control of Fluke Biomedical Testers You can have the medTester 5000C control other instrum
medTester 5000C Operators Manual 11-14
12-1 Chapter 12 Remote Operation In this chapter you find information on both local and remote mode operation of the medTester 5000C. In addition,
medTester 5000C Operators Manual 1-6 Prompts in the autosequence ask you for information that aids in test reporting. The pretest prompts, for example
medTester 5000C Operators Manual 12-2 Additional COM2 Functions Two-way Communications—for commands to and responses from Fluke Biomedical testers and
Biomedical Test System Remote Operation12 12-3 COM2 baud rates are selected from: COM2F2 Note If you have enabled the Wedge adapter, there will
medTester 5000C Operators Manual 12-4 You can connect the optional RS-232 keyboard or laser bar code gun to either serial port. Input from this keyboa
Biomedical Test System Remote Operation12 12-5 Record Output If you enter RECORDS, you see a display similar to this: Configuring Ports for Rec
medTester 5000C Operators Manual 12-6 Configuring Ports for Test Tag Output Configure any of these three ports for test tags as described above for re
Biomedical Test System Remote Operation12 12-7 Forcing A Port An alternative to the above procedure is forcing a port. For example, if you want
medTester 5000C Operators Manual 12-8 UNIT of MEASUREMENT ABBREVIATIONS Abbreviation Unit of Measurement V RMS Volts RMS A RMS Amperes RMS µA RMS
Biomedical Test System Remote Operation12 12-9 ERROR RESPONSES Error Statement Explanation the file has a name embedded into it, the wrong file
medTester 5000C Operators Manual 12-10 Uploading Here is the file upload process: 1. The medTester receives an uploading command from the computer in
Biomedical Test System Remote Operation12 12-11 Downloading The medTester 5000C receives file download commands from the remote computer in the
Biomedical Test System General Information 1 1-7 Checklists are complete sets of preventive maintenance procedures for equipment testing. They c
medTester 5000C Operators Manual 12-12 Remote Commands List Commands sent to the medTester from a remote computer include: • General Administrative C
Biomedical Test System Remote Operation12 12-13 FILE TRANSFER COMMANDS (Note: Usage of these commands is described earlier.) Command Explanat
medTester 5000C Operators Manual 12-14 MEASUREMENT COMMANDS Note: These commands set up a measurement, according to global variables as needed, and
Biomedical Test System Remote Operation12 12-15 MEASUREMENT COMMANDS (µA RMS or DC) LALD ECG leakage current, Left Arm to leads. (µA RMS or DC
medTester 5000C Operators Manual 12-16 WAVE COMMANDS Note: The following commands turn on the specified wave. The wave continues to run until it is
Biomedical Test System Remote Operation12 12-17 Port Diagnostics You can test the input and output of COM1 and COM2, as well as the output of th
medTester 5000C Operators Manual 12-18
13-1 Chapter 13 The Wedge Adapter In this chapter you learn about the Wedge adapter and how to use it, how to attach it to the medTester 5000C, and
medTester 5000C Operators Manual 13-2 Serial Port Expansion By expanding the number of available ports on a medTester 5000C with the Wedge, you have t
Biomedical Test System The Wedge Adapter13 13-3 When you execute a checklist, the medTester 5000C directs any M-type commands through the device
Warranty and Product Support Fluke Biomedical warrants this instrument against defects in materials and workmanship for one full year from the date of
medTester 5000C Operators Manual 1-8 These modules are software enabled for use by the medTester’s firmware. The firmware is held in EPROM (erasable p
medTester 5000C Operators Manual 13-4 Wedge Port Configurations Wedge Port Name (Default Wedge Port No.) Description Non-Invasive Blood Pressure (NIBP
Biomedical Test System The Wedge Adapter13 13-5 Disassembly of the medTester 5000C’s Feet and Tilt Bail Disassemble the medTester 5000C feet an
medTester 5000C Operators Manual 13-6 FROM: Wedge Port (9-pin connection) TO: medTester 5000C (25-pin connection) MEDT COM1 COM1 MEDT COM2 COM2
Biomedical Test System The Wedge Adapter13 13-7 Operating the Wedge Operating the medTester 5000C with the Wedge consists of enabling the wedge,
medTester 5000C Operators Manual 13-8 Configuring the Normal Port The Wedge normal port functions identically to the local input and output modes of t
Biomedical Test System The Wedge Adapter13 13-9 Configuring an Autosequence Port You can execute defibrillator, IV pump, ESU, and SPO2 autoseque
medTester 5000C Operators Manual 13-10 Configure other autosequence ports as described here. Configuring the Checklist Port Whenever you execute a che
Biomedical Test System The Wedge Adapter13 13-11 5. After selecting the normal port, press F5 to store your selection. Changing the Checklist Po
medTester 5000C Operators Manual 13-12 Note Autosequences always communicate at a fixed baud rate; therefore, you don't need to set the rate for
Biomedical Test System The Wedge Adapter13 13-13 5. After setting the baud rate for the port, press F5 to store the setting. Setting Output Po
Biomedical Test System General Information 1 1-9 medTester 5000C Modules medTester 5000C Module Name Functions Enabled medTester Prerequisites
medTester 5000C Operators Manual 13-14
14-1 Chapter 14 ESU Module In this chapter you learn about using the medTester 5000C to control electro-surgery (ESU) autosequence tests with Fluke
medTester 5000C Operators Manual 14-2 Using the ESU Module with the Model 402A and RF303RS ESU Analyzer The ESU module was originally designed to w
Biomedical Test System ESU Module14 14-3 Note Refer to your ESU analyzer operators manual for additional information regarding operation and ran
medTester 5000C Operators Manual 14-4 ESU Autosequence Names You can begin any of the autosequences listed below from the medTester menu commands or d
Biomedical Test System ESU Module14 14-5 Running ESU Autosequence Tests To begin running ESU autosequence tests, connect your medTester 5000C to
medTester 5000C Operators Manual 14-6 To accept previously input prompt information, press YES, and go to the Test Sequence section. If you choose NO,
Biomedical Test System ESU Module14 14-7 8. PHYSICAL INSPECTION—You have two physical inspection lines, each of which can consist of a maximum o
medTester 5000C Operators Manual 14-8 The setup screen prompts the user to set up the ESU for power setting, mode, and output. Push F5 READY when the
Biomedical Test System ESU Module14 14-9 CQM (Contact Quality Monitor) Test The CQM test measures the resistance needed by the ESU device to act
medTester 5000C Operators Manual 1-10 medTester 5000C Modules medTester 5000C Module Name Functions Enabled medTester Prerequisites for Use of Module
medTester 5000C Operators Manual 14-10 ESU Autosequence Descriptions Autosequence Number: A60 Autosequence Name: ASPENX Step and Test Type Load ES
Biomedical Test System ESU Module14 14-11 Autosequence Number: A61 Autosequence Name: VLF30 Step and Test Type Load ESU Mode Output Power
medTester 5000C Operators Manual 14-12 Autosequence Number: A62 Autosequence Name: VLF40 Step and Test Type Load ESU Mode Output Power Setting E
Biomedical Test System ESU Module14 14-13 Autosequence Number: A63 Autosequence Name: VLF4BF Step and Test Type Load ESU Mode Output Power
medTester 5000C Operators Manual 14-14 Autosequence Number: A64 Autosequence Name: VLSSE4 Step and Test Type Load ESU Mode Output) Power Setting
Biomedical Test System ESU Module14 14-15 Customizing ESU Autosequences (For detailed information about customizing autosequences, please refer
medTester 5000C Operators Manual 14-16 Begin by accessing the custom, autosequence, ESU menu as described above. This time, select: MAKEF1 Next, selec
Biomedical Test System ESU Module14 14-17 • Expected Power: Enter the power expected to be generated by the ESU, from 0 to 500 W. • Power hig
medTester 5000C Operators Manual 14-18 STOREF5 Print Your Autosequence Printing any autosequence is accomplished through this menu command: PRINTF3 Us
15-1 Chapter 15 SPO2 Module In this chapter you learn about using the medTester 5000C to control SPO2 autosequence tests with Fluke Biomedical SPO2
Biomedical Test System General Information 1 1-11 Features • Fully automated electrical safety testing • 12–lead ECG/arrhythmia simulation •
medTester 5000C Operators Manual 15-2 PLUS 7), the CardioSat 100 or Index 2XL. All autosequences stored by initialization, customization, or remote do
Biomedical Test System SPO2 Module15 15-3 If SPO2 autosequences are installed, you see this message displayed: The above display format an
medTester 5000C Operators Manual 15-4 SPO2 Autosequence Names You can begin any of the autosequences listed below from the medTester menu commands or
Biomedical Test System SPO2 Module15 15-5 SP02 AUTOSEQUENCES FOR INDEX 2XL Autosequence Name Keyboard Shortcut Command Menu Commands from SP
medTester 5000C Operators Manual 15-6 SP02 AUTOSEQUENCES FOR CARDIOSAT 100 Autosequence Name Keyboard Shortcut Command Menu Commands from SPO2F5AUT
Biomedical Test System SPO2 Module15 15-7 Note The Index 2XL must have been manufactured after 5/21/04 or modified by Fluke Biomedical (ECO 416
medTester 5000C Operators Manual 15-8 Pretest Device Prompts There are eight prompts for information that you can attach to test records. After select
Biomedical Test System SPO2 Module15 15-9 Note See Chapter 5, Autosequences, for a complete explanation of working with prompts. You can turn i
medTester 5000C Operators Manual 15-10 Each autosequence can contain up to ten SPO2 simulation steps, except the INDEX 2XL which can contain up to 15
Biomedical Test System SPO2 Module15 15-11 You should now observe the readings on the pulse oximeter for SPO2 and heart rate. You can enter tes
medTester 5000C Operators Manual 1-12 medTester 5000C Instrument Specifications Line Voltage and Measurements • Hot to Neutral. • Neutral to Ground.
medTester 5000C Operators Manual 15-12 Probe Test For the Oxitest PLUS, there is no probe test built-in. However, it is still possible to include a pr
Biomedical Test System SPO2 Module15 15-13 For the Oxitest PLUS: SPO2 Autosequence Number: A70 Autosequence Name: ADULT1 Settling Time: 25
medTester 5000C Operators Manual 15-14 For the Oxitest PLUS: SPO2 Autosequence Number: A72 Autosequence Name: ADULT3 Settling Time: 25 Beats Heart
Biomedical Test System SPO2 Module15 15-15 For the Oxitest PLUS: SPO2 Autosequence Number: A74 Autosequence Name: NEO2 Settling Time: 25 Bea
medTester 5000C Operators Manual 15-16 For CardioSat 100: SPO2 Autosequence Number: A70 Autosequence Name: NELLCR Manufacturer: Nellcor Settling Ti
Biomedical Test System SPO2 Module15 15-17 For CardioSat 100: SPO2 Autosequence Number: A72 Autosequence Name: HPMRLN Manufacturer: Hewlett
medTester 5000C Operators Manual 15-18 For CardioSat 100: SPO2 Autosequence Number: A74 Autosequence Name: GENERC Manufacturer: Generic–Not Specifi
Biomedical Test System SPO2 Module15 15-19 For the Index 2XL: SPO2 Autosequence Number: A70 Autosequence Name: NELLCR Manufacturer: Nellcor
medTester 5000C Operators Manual 15-20 For the Index 2XL: SPO2 Autosequence Number: A72 Autosequence Name: PMS Manufacturer: PMS M1190 Settling Ti
Biomedical Test System SPO2 Module15 15-21 For the Index 2XL: SPO2 Autosequence Number: A74 Autosequence Name: GENERC Manufacturer: Generic
Biomedical Test System General Information 1 1-13 AAMI Load • Simulated patient load recommended by the Association for the Advancement of Medi
medTester 5000C Operators Manual 15-22 Posttest Prompts When all tests are completed, the medTester 5000C prompts you for: 1. COMMENTS 2. NEXT TEST
Biomedical Test System SPO2 Module15 15-23 Note You can find detailed information about customizing autosequences in Chapter 6, Customize Your
medTester 5000C Operators Manual 15-24 Not Specified BCI Criticare Datascope Datex Hewlett Packard Nellcor Nihon Kohden Novametrix Ohmeda Respironics
Biomedical Test System SPO2 Module15 15-25 Note If you choose "Not Specified", you can keep it generic. However, you must manually se
medTester 5000C Operators Manual 15-26 MOREF4 to select more steps, or ENDF5 to end your selections. 6. STARTING ALARM VALUES: Enter starting value
Biomedical Test System SPO2 Module15 15-27 Print Your Autosequence From the SPO2 custom autosequences menu described above, printing any autose
medTester 5000C Operators Manual 15-28
16-1 Chapter 16 Transcutaneous Pacemaker (Pacer) Module In this chapter you learn about using the medTester 5000C to control transcutaneous pacemake
medTester 5000C Operators Manual 16-2 For related information, see these chapters in this manual: • Chapter 5, Autosequences. • Chapter 6, Customize
Biomedical Test System Transcutaneous Pacemaker (Pacer) Module16 16-3 Pacer Autosequence Names You can begin any of the autosequences listed bel
medTester 5000C Operators Manual 1-14 Toolbox Toolbox allows for the use of external measurement adapters for: • Tachometer—with a range of 100 to 20
medTester 5000C Operators Manual 16-4 Running Pacer Autosequence Tests To begin running pacer autosequence tests, connect your medTester 5000C to your
Biomedical Test System Transcutaneous Pacemaker (Pacer) Module16 16-5 Pretest Device Prompts There are eight prompts for information that can ap
medTester 5000C Operators Manual 16-6 Note See Chapter 5, Autosequences, for a complete explanation of working with prompts. You can turn individual p
Biomedical Test System Transcutaneous Pacemaker (Pacer) Module16 16-7 Pulse Tests For each of the pulse test steps, 1 through 10, the medTester
medTester 5000C Operators Manual 16-8 Demand Test The demand test checks that in the demand mode, the pacer will pulse interactively with ECG waveform
Biomedical Test System Transcutaneous Pacemaker (Pacer) Module16 16-9 Async Test The async test checks that in the asynchronous, or non-demand,
medTester 5000C Operators Manual 16-10 Pacer Autosequence Number: A20 Autosequence Name: LP12P Load: Internal 50 Ohms Step Number Current (in mill
Biomedical Test System Transcutaneous Pacemaker (Pacer) Module16 16-11 Pacer Autosequence Number: A21 Autosequence Name: LP9PM Load: Adapter
medTester 5000C Operators Manual 16-12 Pacer Autosequence Number: A22 Autosequence Name: HPCMPM Load: Adapter Module TQA-12, External 400 Ohms Step
Biomedical Test System Transcutaneous Pacemaker (Pacer) Module16 16-13 Pacer Autosequence Number: A23 Autosequence Name: MDE3PM Load: Adapter
Biomedical Test System General Information 1 1-15 Ground Fault Circuit Interrupter The test receptacle is protected by a ground fault circuit in
medTester 5000C Operators Manual 16-14 Pacer Autosequence Number: A24 Autosequence Name: ZOLLPM Load: Adapter Module TQA-2, External 1000 Ohms Step
Biomedical Test System Transcutaneous Pacemaker (Pacer) Module16 16-15 Customizing Pacer Autosequences (For detailed information about customizi
medTester 5000C Operators Manual 16-16 Begin by accessing the custom, autosequence, pacer menu as described earlier. This time, select: MAKEF1 Next se
Biomedical Test System Transcutaneous Pacemaker (Pacer) Module16 16-17 Reinitialize Factory Default Settings Any customized autosequences can be
medTester 5000C Operators Manual 16-18
17-1 Chapter 17 Non-Invasive Blood Pressure (NIBP) Module In this chapter you learn about using the medTester 5000C to control non-invasive blood pr
medTester 5000C Operators Manual 17-2 For related information, see these chapters in this manual: • Chapter 5, Autosequences. • Chapter 6, Customize
Biomedical Test System Non-Invasive Blood Pressure (NIBP) Module17 17-3 NIBP Autosequence Names You can begin any of the autosequences listed be
medTester 5000C Operators Manual 17-4 Running NIBP Autosequence Tests To begin running NIBP autosequence tests, connect your medTester 5000C to your C
Biomedical Test System Non-Invasive Blood Pressure (NIBP) Module17 17-5 Pretest Device Prompts There are eight prompts for information that can
medTester 5000C Operators Manual 1-16 General Specifications Parameter Specification Power Requirement: 115 V ±10%, 60 Hz only Temperature Ranges:Oper
medTester 5000C Operators Manual 17-6 Note See Chapter 5, Autosequences, for a complete explanation of working with prompts. You can turn individual p
Biomedical Test System Non-Invasive Blood Pressure (NIBP) Module17 17-7 The medTester records the following test parameters from the CuffLink fo
medTester 5000C Operators Manual 17-8 PRESSURE TEST The pressure test allows you to document up to three static pressure levels generated by the CuffL
Biomedical Test System Non-Invasive Blood Pressure (NIBP) Module17 17-9 LEAK TEST The leak test measures the rate of leakage in the NIBP measure
medTester 5000C Operators Manual 17-10 NIBP Autosequence Number: A80 Autosequence Name: ADULT1 Step Number Blood Pressure Heart Rate in BPM (beats
Biomedical Test System Non-Invasive Blood Pressure (NIBP) Module17 17-11 NIBP Autosequence Number: A83 Autosequence Name: NEONT1 Step Number
medTester 5000C Operators Manual 17-12 Customizing NIBP Autosequences (For detailed information about customizing autosequences, please refer to Chapt
Biomedical Test System Non-Invasive Blood Pressure (NIBP) Module17 17-13 Begin by accessing the custom, autosequence, NIBP menu as described ear
medTester 5000C Operators Manual 17-14 After customizing your autosequence, the medTester prompts you to store the autosequence: STOREF5 Print Your Au
Biomedical Test System Index Index-1 Index A AAMI load, 1-13 Accessories, 1-17 Audio Transducer, 2-4 Autosequences defined, 5-1 monitoring envi
Biomedical Test System General Information 1 1-17 Accessories Standard Part # • 20/15–amp Adapter 2195732 • Two Kelvin Cable 2392617 • Two
Index-2 checklists, using, 11-3 Memory checking contents of, 8-6 Menus navigating, 3-4 Modules, list of, 1-8 modules, optional overview, 1-2 Modules,
Biomedical Test System Index Index-3 installing, 13-4 operating, 13-7 setting baud rates, 13-11 setting output ports, 13-13
medTester 5000C Operators Manual Index-4
Notices All Rights Reserved Copyright 2006, Fluke Biomedical. No part of this publication may be reproduced, transmitted, transcribed, stored in a r
medTester 5000C Operators Manual 1-18 Optional (continued) Part # • Printer Cable, medTester to standard PC printer Serial Cable; Male DB25 to Cent
Biomedical Test System General Information 1 1-19 medTester 5000C Module Upgrades Part #Factory Installed • Module 2, Service, RS-232/Printe
medTester 5000C Operators Manual 1-20
2-1 Chapter 2 Installation In this chapter you will find information about the default (factory) settings, power-up sequence, setting the time and d
medTester 5000C Operators Manual 2-2 FACTORY DEFAULT SETTINGS Condition Name Default Setting(s) Output • RECORD and TAG output set to: • COM1–OFF •
Biomedical Test System Installation 2 2-3 medTester 5000C Power-Up Condition Messages Message Condition Indicated LOW BATTERY, SERVICE REQUIRED
medTester 5000C Operators Manual 2-4 4. After entering date and time, insert a thin, non-metallic device in the “clock set enable switch” slot (labele
Biomedical Test System Installation 2 2-5 Enabling Modules If it should become necessary to reinstall modules 2 through 14, perform the steps de
medTester 5000C Operators Manual 2-6 Note If running Windows 95 or 98, with the mouse you can click and hold the Start button, then select Programs an
Biomedical Test System Installation 2 2-7 10. To install more than one module, repeat steps seven (7) through nine (9) for each module which has
Restocking Charges Products returned within 30 days of original purchase are subject to a minimum restocking fee of 15 %. Products returned in excess
medTester 5000C Operators Manual 2-8
3-1 Chapter 3 Instrument Familiarity This chapter familiarizes you with the medTester 5000C. Know Your medTester 5000C The medTester 5000C has two m
medTester 5000C Operators Manual 3-2 MED TESTER 500011413 15 16 17 18 1923851246791011 Figure 3-1. MedTester 5000C Top and Rear Panel Locator
Biomedical Test System Instrument Familiarity 3 3-3 6. ARROW KEYS—Control cursor movement when entering data into the display from the keyboar
medTester 5000C Operators Manual 3-4 16. COM1 PORT—Receives input locally from the optional keyboard, bar code scanning gun, or personal computer or t
Biomedical Test System Instrument Familiarity 3 3-5 The display now shows menu two. Remember that menu two is an extension of menu one. Menu-tw
medTester 5000C Operators Manual 3-6
4-1 Chapter 4 Manual Tests Learn to use the medTester 5000C to manually perform electrical safety tests for line voltage, leakage current, equipment
medTester 5000C Operators Manual 4-2 Hot to Neutral Enter menu commands: MANUALF2VOLTSF1HOT-NEUTF1 Neutral to Ground Enter menu commands: MANUALF2VOLT
Biomedical Test System Manual Tests 4 4-3 External Case Leakage See Figure 4-2, Case External Leakage. 1. Plug the EUT (equipment under test) i
traceable to the National Institute of Standards and Technology (NIST). Devices for which there are no NIST calibration standards are measured against
medTester 5000C Operators Manual 4-4 • POLARITY—normal or reversed, • NEUTRAL—open or closed. The respective condition LEDs light, and the display
Biomedical Test System Manual Tests 4 4-5 • Lead isolation, all leads. • Lead isolation, single lead Note For all lead-to-ground and lead-to-l
medTester 5000C Operators Manual 4-6 RL—right leg, RA—right arm, LA—left arm, LL—left leg, and V—V1 through V6 leads tied together. You can measure V
Biomedical Test System Manual Tests 4 4-7 Isolation Lead Leakage See Figure 4-7, ECG Lead Isolation Test: All Leads and Individual Leads. You c
medTester 5000C Operators Manual 4-8 4. Select the keyboard function key that corresponds to the lead you wish to test. 5. To begin the measurement,
Biomedical Test System Manual Tests 4 4-9 Power Cord Resistance See Figure 4-9, Power Cord Resistance Test. 1. Plug the equipment under test in
medTester 5000C Operators Manual 4-10 External Resistance See Figure 4-10, External Resistance. 1. Connect the dual banana-plug end of one Kelvin cab
Biomedical Test System Manual Tests 4 4-11 Toolbox Toolbox consists of measurements that require the use of three adapters. These adapters, avai
medTester 5000C Operators Manual 4-12 Temperature 1. Plug the temperature adapter test leads into the EXT INPUT jacks on medTester’s front panel. 2.
Biomedical Test System Manual Tests 4 4-13 Figure 4-1. Three Types of Line Voltage Measurements
medTester 5000C Operators Manual 4-14 Figure 4-2. Case External Leakage
Biomedical Test System Manual Tests 4 4-15 Figure 4-3. Case Internal Leakage
medTester 5000C Operators Manual 4-16 Figure 4-4. External Leakage Current
Biomedical Test System Manual Tests 4 4-17 Figure 4-5. ECG Lead Leakage to Ground: All Leads to Ground and Individual Lead to Ground
medTester 5000C Operators Manual 4-18 Figure 4-6. ECG Interlead Leakage
Biomedical Test System Manual Tests 4 4-19 Figure 4-7. ECG Lead Isolation Test: All Leads and Individual Leads
medTester 5000C Operators Manual 4-20 Figure 4-8. Equipment Load Current
Biomedical Test System Manual Tests 4 4-21 Figure 4-9. Power Cord Resistance Test
medTester 5000C Operators Manual 4-22 Figure 4-10. External Resistance
Biomedical Test System Manual Tests 4 4-23 Figure 4-11. Isolated Power System and Ground Fault Test
Biomedical Test System Contents Table of Contents 1: GENERAL INFORMATION SAFETY CONSIDERATIONS ...
medTester 5000C Operators Manual 4-24
5-1 Chapter 5 Autosequences This chapter explains how to use the medTester 5000C’s electrical safety autosequences, two standard monitoring sequence
medTester 5000C Operators Manual 5-2 • Defibrillators—Using Fluke Biomedical defibrillator analyzers Impulse 3000 and Impulse 4000 (see Chapter 9, De
Biomedical Test System Autosequences 5 5-3 Note In the standard shipping configuration, tests are performed both with power turned on and with p
medTester 5000C Operators Manual 5-4 Standard Safety Autosequence Names You can access autosequences from either medTester menu commands or from the k
Biomedical Test System Autosequences 5 5-5 SAFETY AUTOSEQUENCES TEMPMON Temperature Monitor TOOLBOXF2TEMPMONF4 A18 HUMMON Humidity Monitor TOOL
medTester 5000C Operators Manual 5-6 Note All these pre-test prompts can be turned off. However, if using the medTester with an Equipment Management S
Biomedical Test System Autosequences 5 5-7 Autosequence Steps If you have autosequence prompts turned on, you enter the appropriate information.
medTester 5000C Operators Manual 5-8 System Line Voltage Measurements included are: • Hot to neutral, • Neutral to ground, and • Hot to ground. Pow
Biomedical Test System Autosequences 5 5-9 Leakage Current Measurements After the power cord resistance test, the autosequence moves to leakage
Date and Time Setup ... 2-3 Audio Transducer...
medTester 5000C Operators Manual 5-10 ECG Performance Waveforms If you have the Waveform/Extended Testing module installed on your medTester 5000C, an
Biomedical Test System Autosequences 5 5-11 Note See Chapter 7, Performance Waves, for detailed instructions about performance waveform output.
medTester 5000C Operators Manual 5-12 This prompt appears only if you enable the prompt in the CUSTOM menu. See Chapter 6, Customize Your medTester 50
Biomedical Test System Autosequences 5 5-13 Test Tag Printer After performing an autosequence or checklist (test procedure created in your CMMS)
medTester 5000C Operators Manual 5-14 Note To enable or disable any of the five test tag fields (lines), see Test Tag Configuration in Chapter 6, Cus
Biomedical Test System Autosequences 5 5-15 1. From the menu, press F1 (CONFIG) as before to configure your test tag (test tag lines on/off), or
medTester 5000C Operators Manual 5-16 Menu Selection Procedure Both monitoring autosequences are found under AUTOF1. Press the right arrow key on the
Biomedical Test System Autosequences 5 5-17 1. OP CODE—The operator is the person doing the testing. This code can consist of a maximum of three
medTester 5000C Operators Manual 5-18 After you select ENVMONF3 from the menu, medTester prompts— Select MEMORY if you want to store the test record.
Biomedical Test System Autosequences 5 5-19 End-of-Test Prompts At the end of the autosequence, medTester prompts you for the following data: (p
Biomedical Test System Contents What You See in an Autosequence...6-2 How to Customize...
medTester 5000C Operators Manual 5-20 4. For resetting the maximum reading to become the minimum, press the RESET (F1) key. The medTester will prompt
Biomedical Test System Autosequences 5 5-21 Humidity Monitor You can monitor the level of relative humidity over periods of time, print the resu
medTester 5000C Operators Manual 5-22 4. CYCLE TEST TIME IN MINUTES—Enter the time between measurements (1-9 minutes). 5. TRIGGER LEVEL—Enter the tri
6-1 Chapter 6 Customize Your medTester In this chapter you learn how to customize safety autosequences and how to customize the prompt information t
medTester 5000C Operators Manual 6-2 5. After pressing this function key, you can begin scrolling through the steps in the sequence by using your up a
Biomedical Test System Customize Your medTester 6 6-3 You can press Esc to move out of view and back to the previous menu choices, or you can pr
medTester 5000C Operators Manual 6-4 Customizing Menu Steps As you step through the customizing menu, the medTester asks you to enter information. Not
Biomedical Test System Customize Your medTester 6 6-5 LEAKAGE CURRENT TESTS Test Item Legend rev—reverse pol—polarity clsd—closed gnd—ground neu
medTester 5000C Operators Manual 6-6 LEAKAGE CURRENT TESTS Test Item Legend rev—reverse pol—polarity clsd—closed gnd—ground neu—neutral iso—isolated n
Biomedical Test System Customize Your medTester 6 6-7 Customizing a Blank Safety Autosequence If your medTester 5000C contains the Waveform/Exte