Operators Manual
3-4
Alternatively, in the Numeric-Control Mode, press the number keys for a respiration-lead
setting according to Table 3-1, and select RUN:
Table 3-1. Numeric Codes for Respiration-Lead Settings
Respiration-Lead Setting Numeric Code
RESP LEAD LA 389
RESP LEAD LA STORED 390
RESP LEAD LL 391
RESP LEAD LL STORED 392
For baseline impedance, press the number keys for a respiration-baseline setting
according to Table 3-2, and select RUN:
Table 3-2. Numeric Codes for Respiration-Baseline Settings
Respiration-Baseline Setting Numeric Code
RESP BASE 500 OHM 180
RESP BASE 500 STRD 393
RESP BASE 1000 OHM 181
RESP BASE 1000 STRD 394
RESP BASE 1500 OHM 182
RESP BASE 1500 STRD 395
RESP BASE 2000 OHM 183
RESP BASE 2000 STRD 396
Adjusting the Respiration Rate
Respiration rate is measured by the number of times per minute the lungs inflate
(inspiration) and deflate (expiration), expressed as breaths-per-minute (BrPM). Pediatric
respiration rates are two or three times higher than adult rates. Newborn babies, for
example, breathe from forty to fifty times per minute, while adults typically breathe
approximately fifteen times per minute.
The MPS450 simulates nine preconfigured respiration rates (BrPM): 15, 20, 30, 40, 60,
80, 100, and 120, plus a respiration-off condition.
Action in the Menu-Control Mode
1. Press the top-menu key labeled RESP. If the respiration-rate setting has not been
changed, the screen displays the default setting (20 BRPM):
gje024.eps