Fluke-biomedical MPS450 User Manual Page 98

  • Download
  • Add to my manuals
  • Print
  • Page
    / 126
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 97
MPS450
Operators Manual
4-16
Apnea Simulation
Table 4-24 lists the numeric control codes for the apnea simulation functions.
Table 4-24. Numeric Codes for Apnea Simulation Actions
Numeric
Control Code
Action
Remote Control
Entry
152 Start apnea for 12 seconds. A12
153 Start apnea for 22 seconds. A22
154 Start apnea for 32 seconds. A32
150 Turn on apnea continuously. AON
151 Turn off apnea. AOFF
Blood Pressure Functions
The Blood Pressure Functions consists of Blood-Pressure Sensitivity, Blood Pressure
Zeroing, BP Channel 1: Static-Pressure Levels, BP Channel 2: Static-Pressure Levels, BP
Channel 3: Static-Pressure Levels, BP Channel 4: Static-Pressure Levels, BP Channel 1:
Dynamic Waveforms, BP Channel 2: Dynamic Waveforms, BP Channel 3: Dynamic
Waveforms, BP Channel 4: Dynamic Waveforms, BP Channel 1: Respiration Artifact,
BP Channel 2: Respiration Artifact, BP Channel 3: Respiration Artifact, and BP
Channel 4: Respiration Artifact.
Blood-Pressure Sensitivity
Table 4-25 lists the numeric control codes for the blood-pressure sensitivity functions.
Table 4-25. Numeric Codes for Blood-Pressure Sensitivity Actions
Numeric
Control Code
Action
Remote Control
Entry
069 Set blood-pressure sensitivity to 5 μV/V/mmHg. BPSNS5
068 Set Set blood-pressure sensitivity to 40 μV/V/mmHg. BPSNS40
132 Store blood-pressure sensitivity at 5 μV/V/mmHg. (N/A)
133 Store blood-pressure sensitivity to 40 μV/V/mmHg. (N/A)
(N/A) Store the blood-pressure sensitivity BPSNSSTORE
Blood Pressure Zeroing
Table 4-26 lists the numeric control codes for the blood-pressure zeroing functions.
Table 4-26. Numeric Codes for Blood-Pressure Zeroing Actions
Numeric
Control Code
Action
Remote Control
Entry
007 Zero all four blood pressure channels. ZALL
BP Channel 1: Static-Pressure Levels
Table 4-27 lists the numeric control codes for the BP channel 1: static-pressure levels
functions.
Page view 97
1 2 ... 93 94 95 96 97 98 99 100 101 102 103 ... 125 126

Comments to this Manuals

No comments