Pacemaker Waveform
Table 4-8 lists the numeric control codes for the pacemaker waveform functions.
Table 4-8. Numeric Codes for Pacemaker Waveform Actions
Numeric
Control Code
Action
Remote Control
Entry
383 Run atrial pacer wave. ATR
110 Run asynchronous pacer wave. ASN
111 Run demand frequent sinus pacer wave. DFS
112 Run demand occasional sinus pacer wave. DOS
113 Run AV sequential pacer wave. AVS
114 Run noncapture pacer wave. NCA
115 Run nonfunction pacer wave. NFU
Pacemaker Amplitude
Table 4-9 lists the numeric control codes for the pacemaker amplitude functions.
Table 4-9. Numeric Codes for Pacemaker Amplitude Actions
Numeric
Control Code
Action
Remote Control
Entry
384 Set pacer amplitude to 1 mV. PA1
385 Set pacer amplitude to 2 mV. PA2
386 Set pacer amplitude to 5 mV. PA5
242, 387 Set pacer amplitude to 10 mV. PA10
Pacemaker Width
Table 4-10 lists the numeric control codes for the pacemaker width functions.
Table 4-10. Numeric Codes for Pacemaker Width Actions
Numeric
Control Code
Action
Remote Control
Entry
243 Set pacer width to 0.1 ms. PW0.1
244 Set pacer width to 0.5 ms. PW0.5
245 Set pacer width to 1.0 ms. PW1.0
246 Set pacer width to 1.5 ms. PW1.5
247 Set pacer width to 2.0 ms. PW2.0