Table 4-19. Numeric Codes for R-Wave Amplitude Actions (cont.)
Numeric
Control Code
Action
Remote Control
Entry
302 Set R-wave amplitude to 1.00 mV. RWA1.00
303 Set R-wave amplitude to 1.50 mV. RWA1.50
304 Set R-wave amplitude to 2 mV. RWA2.00
305 Set R-wave amplitude to 2.50 mV. RWA2.50
306 Set R-wave amplitude to 3.00 mV. RWA3.00
307 Set R-wave amplitude to 3.50 mV. RWA3.50
308 Set R-wave amplitude to 4.00 mV. RWA4.00
309 Set R-wave amplitude to 4.50 mV. RWA4.50
310 Set R-wave amplitude to 5.00 mV. RWA5.00
311 Set R-wave amplitude to 5.50 mV. RWA5.50
Respiration Functions
The Respiration Functions category consists of Respiration Lead, Respiration Baseline
(Impedance), Respiration Rate, Respiration Amplitude, and Apnea Simulation.
Respiration Lead
Table 4-20 lists the numeric control codes for the respiration lead functions.
Table 4-20. Numeric Codes for Respiration Lead Actions
Numeric
Control Code
Action
Remote Control
Entry
(N/A) Store the respiration lead. RLSTORE
389 Set respiration lead to LA. RLLA
391 Set respiration lead to LL. RLLL
390 Store LA respiration lead. (N/A)
392 Store LL respiration lead. (N/A)
Respiration Baseline (Impedance)
Table 4-21 lists the numeric control codes for the respiration baseline (impedance)
functions.
Table 4-21. Numeric Codes for Respiration Baseline (Impedance) Actions
Numeric
Control Code
Action
Remote Control
Entry
180 Set respiration base to 500 ohms. RB500
181 Set respiration base to 1000 ohms. RB1000
182 Set respiration base to 1500 ohms. RB1500
183 Set respiration base to 2000 ohms. RB2000