601PROXL International Safety Analyzer Operators Manual 2VDE EQUIVALENTPATIENT LEAK-µAVDE EQUIVALENTDEVICE LEAK-µAIEC 1010ACCESSIBLE-V/µA/ INSU
601PRO SERIESXL x Warranty This Warranty is limited and applies only to new products, except for computer-based software, which is covered under a sep
601PRO SERIESXL 5-6 FLUKE BIOMEDICAL 05/15/00 23:48 CONTROL#: PROCEDURE ID: LOCATION: DEVICE TYPE: MANUFACTURER: SERIAL#: ECH
TEST RECORDS 5-7 Fluke Biomedical 601ProXL ELECTRICAL SAFETY TEST DATE: 05/15/01 TIME: 23:48 CONTROL#: PROCEDURE ID: DEVICE TYPE: 6
601PRO SERIESXL 5-8
6-1 Device Records and Templates Using its serial port, the 601PRO can transfer device information to and from a computer, and send stored test resu
601PRO SERIESXL 6-2 1. Connecting the 601PRO and the Host Computer The serial port parameters of the 601PRO and those of the computer’s communication
DEVICE RECORDS 6-3 • At the Transmit Device Information screen, press START. baw123f.eps Transfer of the information should begin immediately. Th
601PRO SERIESXL 6-4 • Load the computer’s communication program and enable it to send a device information file to the 601PRO. Transfer of the inform
DEVICE RECORDS 6-5 4. Device Information Record: Definition of Fields Field Name Maximum Characters CONTROL#: (15)** PROCEDURE ID: (2
601PRO SERIESXL 6-6 ⇒ The Test Standard field must be an exact match with IEC 601-1, VDE 701, VDE 751.1, HEI 95, AAMI, IEC 1010, AS/NZS 3551, or the
DEVICE RECORDS 6-7 6. Deleting Device Records and Templates • From the MAIN MENU, press UTILITIES, then DEVICE RECORDS or TEMPLATES. Press *MORE
NOTICES, WARNINGS AND CONTENT xi Contents Notices ...
601PRO SERIESXL 6-8 • To delete all device records on the 601PRO, press DELETE ALL at the Device Record Utility menu: ☛ Note: Deleting all record
7-1 Testing Devices This chapter provides instructions on how to use the 601PRO to perform routine testing on permanently wired, portable and batte
601PRO SERIESXL 7-2 9. Press the Start Test key to initiate the test. This is a point-to-point resistance measurement. 10. Make sure the reading is b
TESTING DEVICES 7-3 5. Testing Conductive Surfaces 1. Connect the RED test lead to the RED input terminal, and the BLACK test lead to the BLACK in
601PRO SERIESXL 7-4
8-1 Standards and Principles The 601PRO is programmed to use IEC 601-1 as the default test standard. Alternative test standards − VDE 751, VDE 701,
601PRO SERIESXL 8-2 1. Accessing System Setup • From the 601PRO MAIN MENU, press SYSTEM SETUP: baw112f.eps 2. Selecting the Test Standard • At
STANDARDS AND PRINCIPLES 8-3 Manual Operation All manual tests are the same except for Leakage (referred to as Equivalent Device Leakage Current) an
601PRO SERIESXL 8-4 VDE 701 and VDE 751-1 VDE 701 and VDE 751 tests are specified for line voltages of >180V. Earth resistance test current is se
STANDARDS AND PRINCIPLES 8-5 IEC 601-1 Controlled Power Sequence and Test Limits for Auto and Step Tests DC-Only tests will only be performed if DC
601PRO SERIESXL xii Chapter 3: Manual Mode Connecting the Device Under Test ...3-1 The Power-Up Se
601PRO SERIESXL 8-6 IEC 601-1 Controlled Power Sequence and Test Limits for Auto and Step Tests (Continued) IEC 601-1 STANDARD DUT OUTLET CLASS I
STANDARDS AND PRINCIPLES 8-7 IEC 601-1 Conventional Test Sequence and Test Limits for Auto and Step Tests DC-Only tests will only be performed if DC
601PRO SERIESXL 8-8 IEC 601-1 Conventional Test Sequence and Test Limits for Auto and Step Tests (Continued) IEC 601-1 STANDARD DUT OUTLET CLASS I
STANDARDS AND PRINCIPLES 8-9 IEC 601-1 Conventional Test Sequence and Test Limits for Auto and Step Tests (Continued) IEC 601-1 STANDARD CLASS I C
601PRO SERIESXL 8-10 AAMI Controlled Power Sequence and Limits AAMI STANDARD TESTS DUT OUTLET IEC 601 CLASS/TYPE & AAMI CATEGORY* l/B I/BF
STANDARDS AND PRINCIPLES 8-11 IEC 1010 Controlled Power Sequence and Limits (Reflects IEC 61010-1 Second Edition, 2001-2002) IEC 1010 STANDARD CLAS
601PRO SERIESXL 8-12 HEI 95 Controlled Power Sequence and Test Limits for Auto and Step Tests DC-Only tests will only be performed if DC readings ar
STANDARDS AND PRINCIPLES 8-13 HEI 95 Conventional Test Sequence and Test Limits for Auto and Step Tests DC-Only tests will only be performed if DC r
601PRO SERIESXL 8-14 AS/NZS 3551 Controlled Power Sequence and Test Limits for Auto and Step Tests AS/NZS 3551 STANDARD DUT OUTLET CLASS I CLASS
STANDARDS AND PRINCIPLES 8-15 Test Principle: Protective Earth Resistance Notes: S1 = Not applicable S2 = Not applicable S3 = Open S5 = Open b
NOTICES, WARNINGS AND CONTENT xiii Chapter 4: Auto/Step Modes Selecting Auto or Step Mode Testing ...
601PRO SERIESXL 8-16 Test Principle: Earth Leakage Current Notes: S1 = Variable S2 = Variable S3 = Open S4 = Variable S5 = Closed baw136f.eps
STANDARDS AND PRINCIPLES 8-17 Test Principle: Enclosure Leakage Current Notes: S1 = Variable S2 = Variable S3 = Variable S4 = Variable S5 = Clo
601PRO SERIESXL 8-18 Test Principle: Mains Insulation Resistance Notes: S3 = Open baw140f.eps Function: Measures the insulation resistance (at
STANDARDS AND PRINCIPLES 8-19 Test Principle: Applied Part Insulation Resistance Notes: S3 = Open baw141f.eps Function: Measures insulation res
601PRO SERIESXL 8-20 Test Principle: Mains on Applied Part Notes: S1 = Closed S2 = Variable S3 = Closed S4 = Variable S5 = Closed S6 = Variable
STANDARDS AND PRINCIPLES 8-21 Test Principle: Patient Auxiliary Current Notes: S1 = Variable S2 = Variable S3 = Variable S4 = Variable
601PRO SERIESXL 8-22 Test Principle: Patient Leakage Current Notes: S1 = Variable S2 = Variable S3 = Variable S4 = Variable S5 =
STANDARDS AND PRINCIPLES 8-23 Test Principle: VDE, Equivalent Device Leakage Note: S3 = Closed baw145f.eps Function: Measures the curren
601PRO SERIESXL 8-24 Test Principle: VDE, Equivalent Patient Leakage Note: S3 = Closed baw146f.eps Function: Measures the current that
STANDARDS AND PRINCIPLES 8-25 Test Principle: Accessible Voltage Notes: S1 = Variable S2 = Variable S3 = Variable S5 = Closed baw14
601PRO SERIESXL xiv Chapter 8: Standards and Principles Accessing System Setup...
601PRO SERIESXL 8-26 Test Principle: Accessible Leakage Current Notes: S1 = Variable S2 = Variable S3 = Variable S5 = Closed baw148f.eps Functio
9-1 Custom Standards This chapter describes how to define and edit custom test standards. Although the 601PRO has many preprogrammed test standards
601PRO SERIESXL 9-2 • Press EDIT CUSTOM STANDARD: baw115f.eps • At the Select Custom Test Standard screen, choose one of four available test standa
CUSTOM STANDARDS 9-3 • If the test does require limits to be entered, press YES, and then use the numeric keys to enter the limit. The following t
601PRO SERIESXL 9-4 2. Tests Available in a Custom Standard The following table lists all of the tests available in a custom standard, in the order i
CUSTOM STANDARDS 9-5 3. Custom Standard Controlled Power Sequence The following table shows the controlled power sequence for custom test standard
601PRO SERIESXL 9-6 Custom Standard Controlled Sequence (Cont.) IEC-601-1 STANDARD DUT OUTLET Power OFF Delay Reverse OFF Startup Delay Reverse N
CUSTOM STANDARDS 9-7 Custom Standard Controlled Sequence (Cont.) IEC-601-1 STANDARD DUT OUTLET Description Polarity Circuit Line Voltage V N/A
601PRO SERIESXL 9-8 Custom Standard Control Sequence (Cont.) IEC-601-1 STANDARD DUT OUTLET Power OFF Delay Reverse OFF Startup Delay Reverse Norm
10-1 Computer Control The 601PRO can be remotely operated using a host computer. The operator runs all tests normally accessible from the 601PRO&apo
1-1 Introduction and Description 1. Introduction to the 601PRO SeriesXL 2. Accessories 3. Optional Accessories 4. Menu Structure 5. System Ch
601PRO SERIESXL 10-2 Establishing Computer Control Use the checklist that follows to establish computer control of the 601PRO. Refer to the charts
COMPUTER CONTROL 10-3 All command characters are sent in CAPITAL LETTERS (lower-case letters will not be recognized). When a valid command is receiv
601PRO SERIESXL 10-4 Computer Control Commands ENTER DESCRIPTION COMMAND DESCRIPTION NOTES [QUI] Quit COM session Bi-Directional [Q] Quit COM
COMPUTER CONTROL 10-5 Computer Control Commands (Cont.) Voltage Measurement Commands COMMAND DESCRIPTION NOTES [VL1] Voltage L1-EARTH [VL2] Volt
601PRO SERIESXL 10-6 Computer Control Commands (Cont.) Leakage Current Measurement Commands (OUTLET AUTOMATICALLY ON FOR TEST) COMMAND DESCRIPTION
COMPUTER CONTROL 10-7 Computer Control Commands (Cont.) Auxiliary Current Measurement Commands (OUTLET AUTOMATICALLY ON FOR TEST) COMMAND DESCRIPTIO
601PRO SERIESXL 10-8 Computer Control Commands (Cont.) Mains on Applied Part Commands (Cont.* [AMAL] Sink Current ALL-GND AAMI Test Load [AMRA]
COMPUTER CONTROL 10-9 Computer Control Commands (Cont.) Patient Leakage Measurement – DC-Only Commands (OUTLET AUTOMATICALLY ON FOR TEST) COMMAND DE
601PRO SERIESXL 10-10 Computer Control Commands (Cont.) ECG/Performance Wave Commands COMMAND DESCRIPTION NOTES [W01] ECG 30 [W02] ECG 60 [W03
COMPUTER CONTROL 10-11 Computer Control Commands (Cont.) Computer Control Commands Supported for Compatibility to Older Models of the ISA601PROCE CO
601PRO SERIESXL 1-2 Available electrical safety tests include: ♦ Mains Voltage ♦ Dual Lead Voltage ♦ Dual Lead Leakage ♦ Current Consumption ♦ In
601PRO SERIESXL 10-12
11-1 Error Messages, Troubleshooting, and Support The tables in this chapter list error messages that may be received during 601PRO operation, and a
601PRO SERIESXL 11-2 2. Errors and Suggested Corrective Actions Follow the suggested corrective action for errors in the range of 0A00-1900. If the
ERROR MESSAGES AND SUPPPORT 11-3 3. Troubleshooting If possible, please first visit the Fluke Biomedical Technical Assistance Center Web site at ww
601PRO SERIESXL 11-4
A-1 Specifications This appendix details the device specifications for the 601PRO Safety Analyzer. For more information, please contact your Fluke
601PRO SERIESXL A-2 DC Only IEC-601/AAMI IEC-1010 Measurement Range: 0-8,000 µA 0-16,000 µA* Accuracy: ± 1% of reading ± 1µA (@ DC) ±2 % of readi
SPECIFICATIONS A-3 The selectable test current for 1A, 10A, or 25A with open circuit voltage is less than 6 VAC RMS, where open circuit resistance i
601PRO SERIESXL A-4 (c) Specified accuracy for load inductance up to 120 µH and phase angle greater than 10 degrees: 1A and 25 A: 5% of reading ± 30
SPECIFICATIONS A-5 Insulation Resistance Configuration: Measurement taken through the DUT outlet as well as the Patient leads Applied Voltage: 500 V
INTRODUCTION AND DESCRIPTION 1-3 2. Accessories The following accessories are shipped standard with the 601PRO. To order additional quantities, con
601PRO SERIESXL A-6 Serial Port Baud Rate: 2400, 9600, 19,200 Databits: 8 Parity: None Stop bits: 1 Signal levels: Per EI
B-1 Keyboard Options/Barcode Keyboard Wedge An optional keyboard is available for the 601PRO that allow the user complete control of the unit. The e
601PRO SERIESXL B-2 The Fluke Biomedical-specified keyboards do not have a separate numeric keypad but have a second function (Fn) key that makes the
KEYBOARD OPTIONS B-3 2. Attach the Barcode Keyboard Wedge 1. Turn off power to the 601PRO. 2. Connect the barcode keyboard wedge to the keyboard
601PRO SERIESXL B-4 French Keyboard German Keyboard
KEYBOARD OPTIONS B-5 Italian Keyboard English Keyboard
601PRO SERIESXL B-6
C-1 Printer Maintenance The 601PRO printer requires a minimum amount of maintenance. Beyond paper replacement and keeping the printer free of dirt,
601PRO SERIESXL C-2 Paper will automaticallyfeed when inserted.PRINTER HOUSING(inside unit chassis)PAPER PORTROLL CAPPAPER ROLL baw157f.eps Replacing
PRINTER MAINTENANCE C-3 baw156f.eps Replacing the Paper Roll 1. Using the red pull-tab on the printed circuit board (PCB), carefully pull the prin
601PRO SERIESXL 1-4 MAIN MENUTESTS/AUTOMODESCLASS/TYPE UTILITIESSYSTEMSETUP(CHAPTER 2)AUTO(CHAPTER 4)SYSTEMTEST(CHAPTER 3)DEVICERECORDSORTEMPLATES(CH
601PRO SERIESXL C-4
D-1 Test Data Record ASCII Character Formats Test data records must only contain ASCII characters formatted as shown in the table below. The first p
601PRO SERIESXL D-2 Field Description ”351”,”data”,”P/F/N” (Enclosure Leak No L2) ”352”,”data”,”P/F/N” (Enclosure Leak No Earth) ”353”,”dat
ASCII CHARACTER FORMATS D-3 Field Description ”485”,”data”,”P/F/N” (MAP Norm Outlet,Rev Isolation- RA only) ”486”,”data”,”P/F/N” (MAP Norm Ou
601PRO SERIESXL D-4 Field Description ”575”,”data”,”P/F/N” (DC Patient Leak Rev, No Earth- All to Earth) ”580”,”data”,”P/F/N” (DC Patient
ASCII CHARACTER FORMATS D-5 Field Description ”680”,”data”,”P/F/N” (Pat Aux Leak DC Rev No L2 RA-ALL) ”681”,”data”,”P/F/N” (Pat Aux Leak DC Rev
601PRO SERIESXL D-6
INTRODUCTION AND DESCRIPTION 1-5 5. System Characteristics The 601PRO uses a membrane keypad for selection of tests or menu options. The keys are
601PRO SERIESXL 1-6 2VDE EQUIVALENTPATIENT LEAK-µAVDE EQUIVALENTDEVICE LEAK-µAIEC 1010ACCESSIBLE-V/µA/ INSULATION-MVOLTS-VENCLOSURELEAKAGE-µA10CURRENT
INTRODUCTION AND DESCRIPTION 1-7 Keys Used to Enter Device Control Numbers The keys described below (0 through –) can be used to enter test control
601PRO SERIESXL 1-8 ABCDE F baw154f.eps Figure 1-2: 601PRO Front Panel Illustration Front Panel A Applied Part Terminal The jacks allow direct conne
INTRODUCTION AND DESCRIPTION 1-9 DAEB C baw155f.eps Figure 1-3: 601PRO Back Panel Illustration Back Panel A RS232 Connection Allows bi-directional
601PRO SERIESXL 1-10 6. Statement of Compatibility The 601PRO is compatible with and has the same base features as the previous 601PRO safety analyze
2-1 Setting Up the 601PRO This chapter describes the factory default settings and how to change them, and provides instructions for one-time customi
601PRO SERIESXL 2-2 2. Selecting the Test Standard The Select Test Standard option allows the operator to choose among IEC 601-1, VDE 701-1, VD
SETTING UP THE 601PRO 2-3 3. Selecting the Printer Output Before testing, choose whether to use the optional internal printer, an external printer
601PRO SERIESXL 2-4 4. Selecting the RS232 Baud Rate To select the baud rate for the RS232 serial port: • From the 601PRO MAIN MENU, press SY
SETTING UP THE 601PRO 2-5 5. Activating the Beeper To activate/deactivate the beeper: • From the 601PRO MAIN MENU, press SYSTEM SETUP to access t
NOTICES, WARNINGS AND CONTENT iii Notices Fluke Biomedical A Division of Fluke Electronics 6920 Seaway Blvd. Everett, WA 98203 USA Customer Serv
601PRO SERIESXL 2-6 6. Setting the Time and Date The time and date are used to record testing times and their associated dates. The clock can b
SETTING UP THE 601PRO 2-7 To set the date format: • From the 601PRO MAIN MENU, press SYSTEM SETUP to access the Select Setup Function menu. Press
601PRO SERIESXL 2-8 7. Configuring the Enclosure Leakage for the Auto Mode Sequence To take multiple Enclosure Leakage readings when performing
SETTING UP THE 601PRO 2-9 8. Selecting Language Options The Language option allows the operator to select an English, French, German, or Italian e
601PRO SERIESXL 2-10 9. Selecting the DC Option The DC option permits the operator to disable all DC readings, allowing the Auto/Step sequences
SETTING UP THE 601PRO 2-11 10. Selecting the Auto/Step Tests: Controlled Power Sequences or 601CE Conventional Test Sequences The Auto/Step Sequenc
601PRO SERIESXL 2-12 For the controlled power sequences, refer to Chapter 8, Standards and Principles, pages 8-5 and 8-6 for IEC 601 test limits
SETTING UP THE 601PRO 2-13 11. Enabling Stop on Failure The Stop On Failure option allows the operator to stop the execution of an Auto sequence wh
601PRO SERIESXL 2-14 • If YES is selected: During an Auto sequence, when a reading is taken that is outside the limit for the current test, the
SETTING UP THE 601PRO 2-15 12. Configuring for Device Records or Templates1 To speed up the process of entering test variables into the 601 Pro bef
601PRO SERIESXL iv Warnings Use of this instrument is restricted to qualified personnel who recognize shock hazards and are familiar with safety pre
601PRO SERIESXL 2-16 To select between using Device Records or Templates, press SYSTEM SETUP at the MAIN MENU. Next, press *MORE three times at
3-1 Manual Mode This chapter provides a brief overview of 601PRO operation and describes the power-up sequence. It also explains how to set up the 6
601PRO SERIESXL 3-2 2. The Power-Up Sequence baw133f.eps The 601PRO immediately begins a self-check routine at power-up while the following instru
MANUAL MODE 3-3 3. Selecting the Test Standard ☛ Note: Chapter 8, Standards and Principles, provides further instructions for determining and s
601PRO SERIESXL 3-4 4. Selecting the Class/Type The CLASS/TYPE selection sets the Type of EQUIPMENT that is being tested. This affects the limits
MANUAL MODE 3-5 Determine the class/type for the instrument by using the following chart: Class/Type Definitions Class I Protective Earth. (Ther
601PRO SERIESXL 3-6 • At the 601PRO MAIN MENU, press CLASS/TYPE: baw005f.eps Press the SOFT KEY below the desired CLASS/TYPE, or press *MORE for a
MANUAL MODE 3-7 6. Using View Present Settings • The view present settings key allows the operator to inspect or alter the currently selected tes
601PRO SERIESXL 3-8 • Use a SOFT KEY to select a test standard, or press *MORE for further options: baw158f.eps ☛ Note: Altering the test stan
MANUAL MODE 3-9 To choose the current source to be used for the Protective Earth Resistance test, press the view present settings key, and press AM
NOTICES, WARNINGS AND CONTENT v ☛ Exercise extreme caution when a shock hazard is present at the instrument's measurement terminals during th
601PRO SERIESXL 3- 10 CONDITION 3: If at least one Applied Part is defined with a DIFFERENT Type than any of the leads or the EQUIPMENT Type, then t
MANUAL MODE 3-11 4. Make lead type assignments at the Define Lead and Select Type menu: baw039f.eps This selection allows the operator to define
601PRO SERIESXL 3- 12 3. Since LA is Type CF, press TYPE CF. baw042f.eps Press enter to advance to the next lead (LL). 4. Since LL is Type CF,
MANUAL MODE 3-13 7. Manual Operation The 601PRO can be used to perform individual tests manually (Manual mode) without the need to perform an auto
601PRO SERIESXL 3- 14 When in the MAIN MENU or any manual test, the operator may also select a manual test by pressing the test-shortcut keys as ind
MANUAL MODE 3-15 • In tests where the condition of the DUT outlet can be changed, its status is shown on the display and may be modified by pressi
601PRO SERIESXL 3- 16 Shortcut Key 0: Mains Voltage Test and Dual Lead Voltage Test This test can be initiated either from the test-shortcut keys o
MANUAL MODE 3-17 Modify the configuration of the front panel outlet by pressing the appropriate SOFT KEY on the 601PRO: • SOFT KEY 1: Toggles the p
601PRO SERIESXL 3- 18 Shortcut Key 2: Insulation Resistance Test The Insulation Resistance Tests measure (in Megohms) the resistance from DUT outlet
MANUAL MODE 3-19 To perform an Insulation Resistance Applied Part-to-Case test, press AP INSUL while the Insulation Resistance test is displayed. I
601PRO SERIESXL vi Nomenclature Comparable Terminology: International and United States INTERNATIONAL/IEC U.S./AAMI L1 Hot L2 Neu
601PRO SERIESXL 3- 20 Shortcut Key 3: Protective Earth Resistance Test Protective Earth Resistance is measured using the RED test lead attached to t
MANUAL MODE 3-21 ☛ Note: If the calibration fails, the previously stored readings will be used until a passing calibration has occurred. To perf
601PRO SERIESXL 3- 22 To perform a Point-to-Point Resistance test: • Attach the RED lead to the RED input terminal and the BLACK lead to the GREEN
MANUAL MODE 3-23 ⇒ SOFT KEY 1 toggles the DUT outlet Polarity from Normal to Off to Reverse. ⇒ SOFT KEY 2 toggles the DUT outlet from Earth to No
601PRO SERIESXL 3- 24 • Press shortcut key 5. The Enclosure Leakage test appears on the display, and the test begins immediately. baw056f.eps •
MANUAL MODE 3-25 To perform a Patient Leakage Current test using the test-shortcut keys: • From the MAIN MENU, or with the outlet unpowered, plug
601PRO SERIESXL 3- 26 • Press APPLIED PART (SOFT KEY 4) at any time to select the desired applied part leakage current: • Modify the configuratio
MANUAL MODE 3-27 • Select the desired outlet configuration and applied part to test using the appropriate SOFT KEYS: ⇒ SOFT KEY 1 takes one pair o
601PRO SERIESXL 3- 28 Important: If all of the applied parts correspond to the instrument type, the applied parts will be tied together and one rea
MANUAL MODE 3-29 • Press SOFT KEYS 1-4 to select leakage tests: ⇒ SOFT KEY 1 toggles the DUT outlet Polarity from Normal to DUT Off to Reverse Pol
NOTICES, WARNINGS AND CONTENT vii Hazard Warnings ☛ Warning! Power Rating. The 601PRO's mains power input must be connected to a power recep
601PRO SERIESXL 3- 30 Shortcut Key 9: IEC 1010 Accessible Voltage/Leakage Test The Accessible Voltage and Leakage Tests are both accessible using Sh
MANUAL MODE 3-31 baw061f.eps • Press SOFT KEYS 1-3 to select the desired outlet configuration: ⇒ SOFT KEY 1 toggles the DUT outlet Polarity from
601PRO SERIESXL 3- 32 • Press LEAKAGE (SOFT KEY 4). The Accessible Leakage test appears on the display, and the test begins immediately. Display va
MANUAL MODE 3-33 The limit value and the class/type designations will vary depending on the currently selected test standard. The following test co
601PRO SERIESXL 3- 34 WARNING: A 2-beep-per-second signal indicates high voltage present at the applied part terminals while a calibration is being
MANUAL MODE 3-35 To perform a VDE Equivalent Patient Leakage test: • From the MAIN MENU, or with the outlet unpowered, plug the DUT into the 601PR
601PRO SERIESXL 3- 36 Dual Lead Leakage Dual Lead Leakage currents are measured between the RED and BLACK test leads. Measurements may have either a
MANUAL MODE 3-37 ECG Output The 601PRO generates a series of ECG, performance, and arrhythmia waveforms to verify the accuracy of ECG machines/moni
601PRO SERIESXL 3- 38 Sample Waveforms Normal sinus (ECG 30, ECG 60, ECG 120, ECG 180, ECG 240) Atrial Flutter (AFLUT)
MANUAL MODE 3-39 Run (RUN) Atrial Tachycardia Idioventricular (ATACH) (IDIO) R on T (R ON T) PVC 1 (PVC1) Atri
601PRO SERIESXL viii Electromagnetic Interference and Susceptibility ☛ USA FCC CLASS A Warning: Changes or modifications to this unit not expressly a
601PRO SERIESXL 3- 40 Triangle 2 Hz Sine 10, 40, 50, 60, 100 Hz (TRI 2HZ) (SINE 10HZ, SINE 40HZ, SINE 50HZ) (SINE 60 HZ, SI
4-1 Auto/Step Modes This chapter provides an overview of 601PRO Auto and Step mode sequences. The Auto or Step sequence performs all electrical safe
601PRO SERIESXL 4-2 When prompted for a control number or template name, the operator will have several options: 1. Enter a control number or tem
AUTO/STEP MODES 4-3 baw069f.eps To enter a control number for a device record or a template name currently stored in the 601PRO’s database, use a c
601PRO SERIESXL 4-4 • Press YES at the following screen: baw072f.eps The control number or template name is displayed along with the Test Stand
AUTO/STEP MODES 4-5 Protective Earth Resistance during an Auto sequence: baw074f.eps During Step mode operation, the operator must press the NEXT
601PRO SERIESXL 4-6 At the following screen, enter the Technician Name and press ENT. baw078f.eps ☛ Note: If templates was selected during se
AUTO/STEP MODES 4-7 3. Creating/Editing a Device Record or Template ☛ Note: In this section, details about device records are given to explain h
601PRO SERIESXL 4-8 • To edit a device record, press *MORE at the Device Record Utility menu: baw084f.eps • Press EDIT RECORD at the following
AUTO/STEP MODES 4-9 • Select the Class/Type or press *MORE for additional options. baw087f.eps Press ENT. • Select the Applied Current. baw088f.
NOTICES, WARNINGS AND CONTENT ix Based on the testing described below, this instrument bears the CE mark. EC Directive 89/336/EEC Electromagnetic
601PRO SERIESXL 4-10 ☛ Note: When executing the Auto sequence, the operator can press the NEXT key once the DUT is ready to proceed with the te
AUTO/STEP MODES 4-11 *Note: When using the AAMI or Custom test standard, only the applied parts can be assigned; the Type cannot be changed. • Ma
601PRO SERIESXL 4-12 3. Since LA is Type CF, press TYPE CF. baw095f.eps Press ENT to advance to the next lead (LL). 4. Since LL is Type CF,
AUTO/STEP MODES 4-13 Use the test-shortcut keys and the SOFT KEYS to enter information on the following screens. • Enter the Serial Number (Not re
601PRO SERIESXL 4-14 • Enter the Device Manufacturer: baw101f.eps Press ENT. • Enter the Procedure ID: baw102f.eps Press ENT. ☛ Note: Onc
5-1 Test Records Using the Test Record Utility, the operator can choose to transmit, print, or delete all or any one of the result records on board
601PRO SERIESXL 5-2 At the Test Record Utility menu, press TRANSMIT RECORD: baw104f.eps Press a SOFT KEY to enter a Result Number and/or press
TEST RECORDS 5-3 2. Test Data Record: Serial Output Refer to Appendix D for a description of ASCII character formatting. 3. Printing Test Recor
601PRO SERIESXL 5-4 At the Test Record Utility menu, press DELETE RECORDS: baw109f.eps Press YES or NO at the Delete All Test Results? screen: b
TEST RECORDS 5-5 Printing a Header Printing a header before starting a test allows the technician to log important information about the device t