Product Configuration
You can configure the Product through the Ethernet
interface. When a configuration parameter is changed, the
change will be made in the Product and saved on the
micro SD card immediately.
Note
You must install Microsoft Silverlight 5 on Internet
Explorer 7+, Safari 4+, Chrome 12+, or
Firefox 3.6+ to configure the Product through the
internet.
1. Insert a micro SD card that contains the necessary
files into the Product. The SD card must contain the
ClientBin folder that includes ConfigurationWeb.asp
file, the clientaccesspolicy.xml file, and the index.html
file.
Note
The micro SD card must be installed in the
Product if you want to save the configuration. If
you cannot find the micro SD card, talk to your
Fluke Biomedical distributor or call Fluke
Biomedical technical support. See Technical
Support in the front of this manual.
2. Connect the Ethernet port of the Product to a network
or directly to a computer.
3. Push X on the Product to show the Ethernet screen.
4. Push O to select one of the three internet connection
methods: Default, Configured, and DHCP-Client.
The Default selection is the recommended method
when you connect the Product directly to a computer.
The Configured and DHCP-Client selection should
be used when you connect to an existing network.
See the How to Setup an Ethernet Connection section
for instructions to set an IP Address and subnet mask.
The browser-based configuration page in Figure 22
shows in the computer display when an Ethernet
connection is made.
5. To personalize the Product, type a name in the owner
field and a name in the company name field of the web
page.
6. In the upper-left corner of the web page, there are
main menu and submenu hyperlinks that you use to
navigate in the configuration tool.
7. To change Product configuration parameter values,
click on the configuration hyperlink. The configuration
page in Figure 23 shows in the computer display.
The submenu selections are VALUES, CURVES,
TRIGGERS, INTERFACE, and MISC. You click on these
submenu hyperlinks to open the configuration page that
will show the parameters for the selected parameter group.