Gas Flow Analyzer
Accessories and Spare Parts
45
Table 6. Maintenance Tasks
Interval Task
While in
operation
Use the supplied filter.
4 weeks Examine the filter for contamination.
To do this, connect the filter inlet
and outlet to the differential
pressure connection with two T-
pieces. Measure the pressure loss
across the filter with this
connection. The pressure loss for a
flow of 60 l/min cannot be more
than 2 mbar. If the pressure is more
than 2 mbar, the filter must be
replaced.
12 Months Factory calibration to make sure the
Product gives reliable
measurements.