Remote Operations
Codes and Actions 4
4-11
Performance Waves
Table 4-15 lists the numeric control codes for the performance waves functions.
Table 4-15. Numeric Codes for Performance Waves Actions
Numeric
Control Code
Action
Remote Control
Entry
120 Run 2 Hz square performance wave. SQU2
121 Run .125 Hz square performance wave. SQU.125
Table 4-15. Numeric Codes for Performance Waves Actions (cont.)
Numeric
Control Code
Action
Remote Control
Entry
122 Run 2 Hz triangle performance wave. TRI2
123 Run 2.5 Hz triangle performance wave. TRI2.5
124 Run 30 BPM pulse performance wave. PUL30
125 Run 60 BPM pulse performance wave. PUL60
207 Run 0.5 Hz sine performance wave. SIN0.5
208 Run 5 Hz sine performance wave. SIN5
209 Run 10 Hz sine performance wave. SIN10
210 Run 40 Hz sine performance wave. SIN40
211 Run 50 Hz sine performance wave. SIN50
212 Run 60 Hz sine performance wave. SIN60
213 Run 100 Hz sine performance wave. SIN100
Performance Wave Amplitude
Table 4-16 lists the numeric control codes for the performance wave amplitude functions.
Table 4-16. Numeric Codes for Performance Wave Amplitude Actions
Numeric
Control Code
Action
Remote Control
Entry
272 Set performance amplitude to 0.05 mV. PFA0.05
273 Set performance amplitude to 0.10 mV. PFA0.10
274 Set performance amplitude to 0.15 mV. PFA0.15
275 Set performance amplitude to 0.20 mV. PFA0.20
276 Set performance amplitude to 0.25 mV. PFA0.25
277 Set performance amplitude to 0.30 mV. PFA0.30
278 Set performance amplitude to 0.35 mV. PFA0.35
279 Set performance amplitude to 0.40 mV. PFA0.40
280 Set performance amplitude to 0.45 mV. PFA0.45