Introduction and Specifications
Viewing Current MPS450 Parameters 1
1-11
Table 1-4. Parameter Viewing List
View
Page
Wave/Function Current Parameters Displayed
ECG NSR NSR rate, patient type, ST-segment elevation, NSR amplitude,
and ECG artifact.
ECG performance Performance wave and amplitude
ECG R-wave detection R-wave rate, width, and amplitude
ECG Arrhythmia Arrhythmia and ECG artifact
1
Fetal/maternal IUP wave, IUP period, and fetal heart rate
2 Respiration Respiration rate, amplitude, lead, and baseline
3 Blood Pressure BP sensitivity; and settings BP1, BP2, BP3, BP4
4
Temperature/Cardiac Output Temperature, thermistor, cardiac-output wave, and
CO-injectate temperature and flow
2. Select < or > to cycle through the other three VIEW pages, which display as screens
similar to the following:
gje004.eps
gje005.eps
gje006.eps
3. To exit viewing, press the VIEW key again, or press the ESC key to return to the
previous control mode.
4. While viewing settings, the MODE key is inactive. Press the ESC key to exit
viewing before changing modes.