Cardiac Functions
ECG Testing 2
2-25
screen is active when displayed and remains active until the setting is changed.
Alternatively, in the Numeric-Control Mode, press the number keys for R-wave rate
according to Table 2-19, and select RUN:
Table 2-19. Numeric Codes for R-Wave-Rate Settings
R-Wave-Rate Setting Numeric Code
30 BPM 312
60 BPM 313
80 BPM 314
120 BPM 315
200 BPM 316
250 BPM 317
Setting R-Wave Width
The MPS450 offers twenty-two preconfigured settings (ms) for R-wave width: nineteen
settings ranging from 20 to 200 ms (10 ms steps), plus three settings at 8, 10, or 12 ms.
Action in the Menu-Control Mode
1. Press the top-menu key labeled PERF.
2. Select RWDET.
3. Select SEL to toggle to the screen for adjusting R-wave width (WIDTH/ampl), with
WIDTH in upper-case letters.
4. Select DOWN or UP to scroll to the desired R-wave width.
5. The width identified on the LCD screen is active when displayed and remains active
until the setting is changed.
Alternatively, in the Numeric-Control Mode, press the number keys for R-wave width
according to Table 2-20, and select RUN:
Table 2-20. Numeric Code R-Wave Width Settings
R-Wave-Width Setting Numeric Code
8 ms 318
10 ms 319
12 ms 320
20 ms 321
30 ms 322
40 ms 323
50 ms 324
60 ms 325
70 ms 326
80 ms 327
90 ms 328