5. Select DOWN or UP to scroll to a different static-level setting.
6. Select RUN. On the LCD screen, RUN flashes to indicate the setting is active. This
value remains active until the setting is changed.
7. Press the ESC key to return to the previous screen.
8. Press the ESC key again to return to the top menu BLOOD PRESSURE.
Alternatively, in the Numeric-Control Mode, press the number keys for a static-pressure-
level setting according to Table 2-23, and select RUN:
Table 2-23. Numeric Codes for Static-Pressure-Level Settings
Static-Pressure-Level Setting Numeric Code
BP1 -10 mmHg 342
BP1 Zero 343
BP1 80 mmHg 344
BP1 160 mmHg 345
BP1 240 mmHg 346
BP1 320 mmHg 347
BP1 400 mmHg 348
BP2 -10 mmHg 351
BP2 Zero 352
BP2 50 mmHg 353
BP2 100 mmHg 354
BP2 150 mmHg 355
BP2 200 mmHg 356
BP2 240 mmHg 357
BP3 -5 mmHg 360
BP3 Zero 361
BP3 20 mmHg 362
BP3 40 mmHg 363
BP3 60 mmHg 364
BP3 80 mmHg 365
BP3 100 mmHg 366
BP4 -5 mmHg 369
BP4 Zero 370
BP4 20 mmHg 371
BP4 40 mmHg 372
BP4 60 mmHg 373
BP4 80 mmHg 374
BP4 100 mmHg 375