Remote Operations
Codes and Actions 4
4-17
Table 4-27. Numeric Codes for BP Channel 1: Static-Pressure Levels Actions
Numeric
Control Code
Action
Remote Control
Entry
342 Set BP1 to -10 mmHg static. P1S-10
343 Set BP1 to zero. P1S0
344 Set BP1 to 80 mmHg static. P1S80
345 Set BP1 to 160 mmHg static. P1S160
346 Set BP1 to 240 mmHg static. P1S240
347 Set BP1 to 320 mmHg static. P1S320
348 Set BP1 to 400 mmHg static. P1S400
BP Channel 2: Static-Pressure Levels
Table 4-28 lists the numeric control codes for the BP channel 2: static-pressure levels
functions.
Table 4-28. Numeric Codes for BP Channel 2: Static-Pressure Levels Actions
Numeric
Control Code
Action
Remote Control
Entry
351 Set BP2 to -10 mmHg static. P2S-10
352 Set BP2 to zero. P2S0
353 Set BP2 to 50 mmHg static. P2S50
354 Set BP2 to 100 mmHg static. P2S100
355 Set BP2 to 150 mmHg static. P2S150
356 Set BP2 to 200 mmHg static. P2S200
357 Set BP2 to 240 mmHg static. P2S240
BP Channel 3: Static-Pressure Levels
Table 4-29 lists the numeric control codes for the BP channel 3: static-pressure levels
functions.
Table 4-29. Numeric Codes for BP Channel 3: Static-Pressure Levels Actions
Numeric
Control Code
Action
Remote Control
Entry
360 Set BP3 to -5 mmHg static. P3S-5
361 Set BP3 to zero. P3S0
362 Set BP3 to 20 mmHg static. P3S20
363 Set BP3 to 40 mmHg static. P3S40
Page view 98