Alternatively, in the Numeric-Control Mode, press the number keys for a dynamic-wave
setting according to Table 2-25, and select RUN:
Table 2-25. Numeric Codes for Dynamic-Wave Settings
Dynamic-Wave Setting NumericCode
BP1 Arterial 120/80 060
BP1 Radial Art 120/80 061
BP1 Left Vent 120/00 062
BP1 Right Vent 25/00 063
BP2 Arterial 120/80 070
BP2 Radial Art 120/80 071
BP2 Left Vent 120/00 072
BP2 Right Vent 25/00 073
BP2 Pulmonary Art 25/10 074
BP2 Pulm Art Wedge 10/2 075
BP2 Left Atrium 14/4 076
BP2 Right Atrium CVP 15/10 077
BP3 Arterial 120/80 080
BP3 Radial Art 120/80 081
BP3 Left Vent 120/00 082
BP3 Right Atrium CVP 15/10 083
BP3 Pulmonary Art 25/10 084
BP3 Pulm Art Wedge 10/2 085
BP3 Left Atrium 14/4 086
BP3 Right Vent 25/00 087
Adding Respiration Artifact to the BP Signal
The MPS450 adds respiration artifact to any BP-channel dynamic wave, including the
Swan-Ganz-wave series. The artifact is turned on separately for each of the four BP
channels.
The following peak-to-peak changes (mmHg) are caused by the respiration wave, with
values independent of the programmed respiration amplitude: