Cardiac Functions
Blood Pressure Function 2
2-33
for BP channel 1, a delta change from 0 to 16;
for BP channel 2, a delta change from 0 to 12;
for BP channel 3, a delta change from 0 to 5; and
for BP channel 4, a delta change from 0 to 3.
Action in the Menu-Control Mode
1. Press the top-menu key labeled BP.
2. Select BP1, BP2, BP3, or BP4 to display the channel’s screen.
3. Select DYNA to display the DYNAMIC WAVE screen for the selected channel.
4. Select ARTF. On the LCD screen, OFF toggles to ON.
5. Select RUN. On the LCD screen, RUN flashes to indicate the identified waveform is
running with the artifact applied.
6. Press the ESC key to return to the selected BP channel screen. The LCD screen
identifies the waveform that is running and indicates if the artifact is active (ARTF
ON):
gje032.eps
7. Press the ESC key again to return to the top menu BLOOD PRESSURE. The LCD
screen indicates that the selected waveform (in the following example, 120/80) on the
selected channel (2) is running with a respiration artifact (A prefixed to 120/80):
gje033.eps
Alternatively, in the Numeric-Control Mode, press the number keys for a respiration-
artifact setting according to Table 2-26, and select RUN:
Table 2-26. Numeric Codes for Respiration-Artifact Settings
Respiration-Artifact Setting Numeric Code
BP1 Respiration Artifact Off 349
BP1 Respiration Artifact On 350
BP2 Respiration Artifact Off 358
BP2 Respiration Artifact On 359
BP3 Respiration Artifact Off 367